NABERTHERM L1/12 전기로 (1L, 1200°C)
생산업체: NABERTHERM Model: L1/12 - 견적요구
- 연락처
Controler: Option
Tmax : 1200 °C
Inner dimensions in: 90x115x110 mm
Volume: 1 Liters
Outer dimensions in: 250x265x340 mm
Input: 220V-1pha
Load: 1.5 kw
Weight: 10 kg
Time to Tmax: 25 min
Muffle Furnaces with Flap Door
Cataloge Nabertherm Laboratory
- Option Code L-030K1EN with controler P330 :
+ Programmable power outlet
+ Real-time clock
+ Interface for MV software
Datasheet controler Nabertherm Laboratory
L 1/12 | L 5/11 |
Our L 1/12 - LT 40/12 product line is the right choice for daily laboratory use. These models stand out for their excellent workmanship, advanced and attractive design, and high level of reliability. The furnaces come equipped with either a flap door or lift door at no extra charge.
- Tmax 1100 °C or 1200 °C
- Heating from two sides by ceramic heating plates (heating from three sides for models L 24/11 - LT 40/12)
- Ceramic heating plates with integral heating element which is safeguarded against fumes and splashing, and easy to replace
- Highly durable cured vacuum fiber module lining
- Housing made of sheets of textured stainless steel
- Dual shell housing for low external temperatures and high stability
- Optional flap door (L) which can be used as work platform or lift door (LT) with hot surface facing away from the operator
- Adjustable air inlet integrated in door (see illustration)
- Exhaust air outlet in rear wall of furnace
- Solid state relays provide for low-noise operation
Additional equipment
- Chimney, chimney with fan or catalytic converter (not for L 1)
- Over-temperature limiter with manual reset for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the oven and load
- Protective gas connection to purge with non-flammable protective or reaction gases
- Manual or automatic gas supply system
- Process control and documentation with Controltherm MV software package
Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Connected | Electrical | Weight | Minutes | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Flap door | °C | w | d | h | in l | W | D | H | load kW | connection* | in kg | to Tmax² |
L 3/11 | 1100 | 160 | 140 | 100 | 3 | 380 | 370 | 420 | 1.2 | 1-phase | 20 | 60 |
L 5/11 | 1100 | 200 | 170 | 130 | 5 | 440 | 470 | 520 | 2.4 | 1-phase | 35 | 60 |
L 9/11 | 1100 | 230 | 240 | 170 | 9 | 480 | 550 | 570 | 3.0 | 1-phase | 45 | 75 |
L 15/11 | 1100 | 230 | 340 | 170 | 15 | 480 | 650 | 570 | 3.5 | 1-phase | 55 | 90 |
L 24/11 | 1100 | 280 | 340 | 250 | 24 | 560 | 660 | 650 | 4.5 | 3-phase | 75 | 95 |
L 40/11 | 1100 | 320 | 490 | 250 | 40 | 600 | 790 | 650 | 6.0 | 3-phase | 95 | 95 |
L 1/12 | 1200 | 90 | 115 | 110 | 1 | 250 | 265 | 340 | 1.5 | 1-phase | 10 | 25 |
L 3/12 | 1200 | 160 | 140 | 100 | 3 | 380 | 370 | 420 | 1.2 | 1-phase | 20 | 75 |
L 5/12 | 1200 | 200 | 170 | 130 | 5 | 440 | 470 | 520 | 2.4 | 1-phase | 35 | 75 |
L 9/12 | 1200 | 230 | 240 | 170 | 9 | 480 | 550 | 570 | 3.0 | 1-phase | 45 | 90 |
L 15/12 | 1200 | 230 | 340 | 170 | 15 | 480 | 650 | 570 | 3.5 | 1-phase | 55 | 105 |
L 24/12 | 1200 | 280 | 340 | 250 | 24 | 560 | 660 | 650 | 4.5 | 3-phase | 75 | 110 |
L 40/12 | 1200 | 320 | 490 | 250 | 40 | 600 | 790 | 650 | 6.0 | 3-phase | 95 | 110 |
Model | Tmax | Inner dimensions in mm | Volume | Outer dimensions in mm | Connected | Electrical | Weight | Minutes | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Lift door | °C | w | d | h | in l | W | D | H¹ | load kW | connection* | in kg | to Tmax² |
LT 3/11 | 1100 | 160 | 140 | 100 | 3 | 380 | 370 | 420+165 | 1.2 | 1-phase | 20 | 60 |
LT 5/11 | 1100 | 200 | 170 | 130 | 5 | 440 | 470 | 520+220 | 2.4 | 1-phase | 35 | 60 |
LT 9/11 | 1100 | 230 | 240 | 170 | 9 | 480 | 550 | 570+290 | 3.0 | 1-phase | 45 | 75 |
LT 15/11 | 1100 | 230 | 340 | 170 | 15 | 480 | 650 | 570+290 | 3.5 | 1-phase | 55 | 90 |
LT 24/11 | 1100 | 280 | 340 | 250 | 24 | 560 | 660 | 650+335 | 4.5 | 3-phase | 75 | 95 |
LT 40/11 | 1100 | 320 | 490 | 250 | 40 | 600 | 790 | 650+335 | 6.0 | 3-phase | 95 | 95 |
LT 3/12 | 1200 | 160 | 140 | 100 | 3 | 380 | 370 | 420+165 | 1.2 | 1-phase | 20 | 75 |
LT 5/12 | 1200 | 200 | 170 | 130 | 5 | 440 | 470 | 520+220 | 2.4 | 1-phase | 35 | 75 |
LT 9/12 | 1200 | 230 | 240 | 170 | 9 | 480 | 550 | 570+290 | 3.0 | 1-phase | 45 | 90 |
LT 15/12 | 1200 | 230 | 340 | 170 | 15 | 480 | 650 | 570+290 | 3.5 | 1-phase | 55 | 105 |
LT 24/12 | 1200 | 280 | 340 | 250 | 24 | 560 | 660 | 650+335 | 4.5 | 3-phase | 75 | 110 |
LT 40/12 | 1200 | 320 | 490 | 250 | 40 | 600 | 790 | 650+335 | 6.0 | 3-phase | 95 | 110 |
¹Including opened lift door *Please see page 60 for more information about supply voltage ²If connected at 230 V 1/N/PE rsp. 400 V 3/N/PE |