Measuring range Setting range of standard value Measuring current Accuracy
SLOW FAST
100mΩ 5mΩ~ 100mΩ 100mA within ±0.02%±2α±2d within ±0.03%±3α±2±[2/(1+n)] d
1Ω 100.1mΩ~ 1.00Ω 100mA within ±0.02%±α±1d within ±0.02%±α±2d±[2/(1+n)] d
10Ω 1.001Ω~ 10.0Ω 50mA within ±0.02%±α±1d within ±0.02%±α±2d±[2/(1+n)] d
100Ω 10.01Ω~ 100Ω 10mA within ±0.02%±1d within ±0.02%±2d±[1/(1+n)] d
1kΩ 100.1Ω~ 1kΩ 5mA within ±0.02%±1d within ±0.02%±2d±[1/(1+n)] d
10kΩ 1.001kΩ~ 10.0kΩ 0.5mA within ±0.02%±1d within ±0.02%±2d±[1/(1+n)] d
100kΩ 10.01kΩ~ 100kΩ 50μA within ±0.02%±1d within ±0.02%±2d±[1/(1+n)] d
1MΩ 100.1kΩ~ 1.00MΩ 5μA within ±0.02%±1d within ±0.05%±2d±[1/(1+n)] d
10MΩ 1.001MΩ~ 10.0MΩ 0.5μA within ±0.03%±1d within ±0.2%±4d±[1/(1+n)] d
100MΩ 10.01MΩ~ 109MΩ 0.05μA within ±0.1%±2d ----
Measuring time Remote start Free running
SLOW FAST SLOW FAST
+about 18msec. about 0.7msec. about 30 times/sec. about 60 times/sec.
~about 400msec. ~about 400msec. ~about 25 times/sec. ~ about 50 times/sec.
EOC [End of comparison] pulse width: 1~250msec. or continuative
Measuring method: 2 or 4-terminal measurement
Setting range for judgment value: % Measurement:±99.99% ; Absolute measurements:00000~20000
Operation condition [Temp.]: 5℃~+40℃ [Humidity] less than 85%
Power supply: AC85V~265V, 50~60Hz, about 50VA
Outer dimension: about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.)
Weight: about 3.5kg