直接联系认证专家
Refractive index [nD]
Range: 1.26 nD to 1.72 nD
Max. resolution: 0.000001 nD
Accuracy¹: ±0.00002 nD
Repeatability: ±0.0000015 nD
Brix
Range: 0 °Brix to 100 °Brix
Display resolution: 0.001 °Brix
Accuracy: ±0.015 °Brix
Temperature
Temperature range: 4² °C to 105 °C
Temperature probe accuracy¹: ±0.03 °C
Temperature probe stability: ±0.002 °C
Display: 7" LCD touch screen
Communication interfaces:
1 x Ethernet
3 x USB B
CAN-Bus for connection with DMA density meter, MCP polarimeter, SVM viscometer, Xsample 520 sample changer
Features: User management, sample-specific configurations, pass/fail, preconfigured methods, AP Connect ready
Data storage: 10,000 measuring results, 400 products
General
Min. sample volume: 70 uL
Max. allowed pressure in micro flow cell: 10 bar
Ambient temperature: 5 °C to 40 °C
Materials in contact with the sample
Prism: Synthetic sapphire
Sample well: Stainless steel, optionally Ni-alloy
Seal: FFKM (perfluoroelastomer)
Compounds
Light source: LED light source, average lifetime > 100,000 hours
Wavelength: 589.3 nm (by wavelength-adjusted interference filter)
Power requirements: 100-240 VAC +10 %/-15 %, 50/60 Hz, min. 10 W, max. 100 W, depending on sample temperature setting and ambient temperature
Dimensions W x H x D: 225 mm x 115 mm x 330 mm
Weight: 3.5 kg
¹: Valid at standard conditions (T=20 °C, ambient temperature 20 °C to 25 °C
²: At max. ambient temperature of 35 °C