For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Anton Paar DMA 4200 M 密度计 (0 g/cm³ ~ 3 g/cm³, -10°C - 200°C)

联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Measuring range

Density: 0 g/cm³ to 3 g/cm³

Temperature: -10 °C to 200 °C (14 °F to 392 °F)

Pressure range: 0 bar to 500 bar (7250 psi)

Accuracy*

Density: 0.0002 g/cm³

Temperature: 0.03 °C (0.05 °F)

Repeatability, s.d.**

Density: 0.00005 g/cm³

Temperature: 0.01 °C (0.02 °F)

Reproducibility, s.d.**

Density: 0.0001 g/cm³

Digital Resolution: 0.00001 g/cm³

FillingCheck™: Yes (at ambient pressure and up to 150 °C)

Full range viscosity correction: Yes (at ambient pressure)

Temperfect™: Yes (at ambient pressure)

Pulsed Excitation Method: Yes

Minimum sample volume: 2 mL

Wetted materials: Hastelloy C276

Footprint: 510 mm x 330 mm (20 in x 13 in)

ASTM standards: D4052, D5002

Power supply: AC 100 to 240 V, 50 to 60 Hz, 250 VA

Connections: Pressure fittings for metal tubes with 1/8" outer diameter

Interfaces: 4 x USB (2.0 full speed), 1 x Ethernet (100 Mbit), 1 x CAN Bus, 1 x RS-232, 1 x VGA

*this is valid under ideal measuring and sample conditions only (at ambient pressure)

**this is valid under ideal measuring and sample conditions only (at ambient pressure), according to ISO 5725