Detector: Pyroelectric DLaTGS detector

Optics: Hermetically sealed aluminum casing with gold-coated mirrors, KBr windows and beamsplitter

Signal-to-noise ratio: 55,000:1 (1 min, 8 cm⁻¹, 2,100 cm⁻¹ to 2,200 cm⁻¹)

Spectral range: 350 cm⁻¹ to 7,500 cm⁻¹

Spectral resolution: 1.4 cm⁻¹ to 16 cm⁻¹

Wavenumber accuracy: <0.05 cm⁻¹ at 900 cm⁻¹ to 3,000 cm⁻¹

Wavenumber precision: Repeatability <0.0005 cm⁻¹ at 2,000 cm⁻¹ (standard deviation of 10 repeated measurements)

Photometric accuracy: Better than 0.06 % transmittance

Typical measurement time: <30 seconds

Laser type: Single mode Vertical Cavity Surface Emitting Laser (VCSEL)

Laser class: Class 1, enclosed hermetically

IR source: SiC composite

Interferometer: Permanently aligned cube corner interferometer

Desiccant: Molecular sieve with color indicator, user-replaceable

Operating temperature range: 10 °C to 30 °C (50 °F to 86 °F) (non-condensing)

Instrument dimensions (L x W x H): 365 x 315 x 204 mm (14.4 in x 12.4 in x 8.0 in)

Cell compartment width: 152 mm

Weight: 11.7 kg (25.8 lbs)

Power supply: AC 100 V to 240 V, 47 Hz to 63 Hz; DC 24 V, max. 3.75 A

Typical power consumption: 24 W (during operation)

Communication interfaces: 4 x USB 2.0 / CAN / Ethernet

Data export formats: .csv, .pdf, .spc

Data management: AP Connect

Display: No touchscreen

Controls: AP Spectroscopy Suite required