Viscosity range

Operation with internal air pump

Operation with 2 bar external compressed air: 

12.000 mPa·s at ambient and measuring temperature

36.000 mPa·s at ambient and measuring temperature

Filling mode: 

Pressure mode

with integrated piston pump

with air pump of master instrument

with compressed air

Suction mode (optional)

Minimum sample volume per measurement: approx. 3 mL (with master instrument DMA™ M)

depending on system configuration

Sample recovery: optional

Rinse consumption per cycle: approx. 4 mL

Sample throughput: max. 9 per hour

depending on application and used pressure source

Number of sample vials: 71 positions for 12 ml vials

optional: 35 positions for 40 mL or 35 positions for 20 mL vials

customer specific magazines on request

Automatic filling and drying: Piston pump, internal air pump of master instrument or compressed air selectable

Compressed air supply 

(optional): 1.80 bar – 2.20 bar (26 psi – 32 psi) relative external compressed air

typically: 2.0 bar (29 psi) rel.

Necessary quality of air: Class 3 from ISO 8573-1 

max. particle size: 5 μm

max. pressure dew point:-20°C (-4°F)

max. oil content: 1 mg/m3

Environmental conditions (EN 61010): Indoor use only

Air humidity: 10 to 90% relative humidity, non-condensing

Operating temperature: +15 to +35 °C (+59 to +95 °F)

Power supply: supplied by master instrument

Mains voltage according to the type plate at the rear of master instrument (protective grounding required)

Dimensions excl. master instrument (L x W x H): 370 mm x 490 mm x 430 mm (14.6 in x 19.3 in x 17 in)

Weight excl. master instrument: approx. 48 kg (105.8 lbs)

Optional bar code reader: Laser Line Scanner, 1D Code Reader class 2