The PRM-TANKα(FLN) Process Refractometer features a new design of the detection unit, which is to be mounted on the wall of a vessel. It is ideal for a variety of vessel types, such as distillers, condensers, blenders, fermenters, extractors, and filtration tanks. The flat sample stage prevents sample buildups, keeping the prism clean over an extended period of time.

Measurement item: Refractive Index (nD), Brix,concentration, temperature

Range

Refractive index (nD) : 1.31700 to 1.51000

Brix : 0.00 to 85.00%

Resolution

Refractive index (nD):0.0001 or 0.00001

Brix : 0.1 or 0.01%

Accuracy

Refractive index (nD) : ±0.0001

Brix : ±0.1%

Temperature compensation: 5.0 to 100.0゚C

Output: RS-232C, DC4 to 20mA

Alarm output: Open-collector output for high and low limit settings (alarm output)

Cable: Cable between the detection and display units : 15m standard (extension available up to 200m)