Measure range: 0~10/100/1000ppm/1%/25.00% O2
Accuracy:
0.01~9.99ppm O2:±5% FS
10.0~99.9ppm O2:±3% FS
100~1000ppm O2:±2% FS
Repeatability:
0.01~9.99ppm O2:±2.5% FS
10.0~99.9ppm O2:±1.5% FS
Stability:
100~1000ppm O2:±1% FS
0.01~9.99ppm O2:±2.5% FS/7d
10.0~99.9ppm O2:±1.5% FS/7d
100~1000ppm O2:±1% FS/7d
Response time: T90<60s(25°C)
Recovery time: 60 minutes,from air (20.94%) to 10 ppm
Sensor: Fuel cell oxygen sensor
Measuring media: nitrogen,hydrogen,helium,argon and hydrocarbon gases
Charging power supply: 220VAC±10%,50/60Hz,when transformer is configured,And 85~264VAC,50/60Hz,when switch power supply is configured.
Environment temperature: -10~45°C Environment Humidity:<80%RH atmospheric conditions±10% (Gas outlet should be atmospheric conditions)
Analyzer size: L×W×H=244mm×258.5mm×147.5mm