For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Chauvin Arnoux C.A 1623 热电阻温度校准仪 (7 loại đầu đo)

ModelC.A 1623
P/NP01654623
原产地China
保修1 年
不含增值税
31,740,485 VND
含增值税
34,279,724 VND
安全结算
品质保证
轻松更换与退货
支持配送

- Range: 0.00 Ω - 400.00 Ω; 400.0 Ω - 1,500.0 Ω;  1,500.0 Ω - 3,200.0 Ω
- Accuracy 4-wire input/ 2-wire/3-wire input/ Output: ± 0.1 Ω ; ± 0.5Ω; ±1Ω/±2Ω
- Acceptable excitation in: 0.1 - 0.5 mA/ 0.5mA - 3.0mA ; 0.05mA - 0.8mA; 0.05mA - 0.8mA
- Pt10 385
+ Range: - 200 to + 800 °C 
+ Acceptable excitation in 0.1 - 3.0 mA
- Pt50 385 
+ Range: - 200 to + 800 °C 
+ Accuracy 4-wire input/ 2-wire/3-wire input/ Output: 0.7°C / 1.0°C / 0.7°C 
+ Acceptable excitation in 0.1 - 3.0 mA
- Pt100 385
+ Range: - 200 to + 800 °C   
+ Accuracy 4-wire input/ 2-wire/3-wire input/ Output: 0.33°C / 0.5°C / 0.33 °C
+ Acceptable excitation in 0.1 - 3.0 mA
- Pt200 385
+ Range:  200 to + 250 °C / +250 to + 630 °C 
+ Accuracy 4-wire input/ 2-wire/3-wire input/ Output: 0.2 °C/ 0.3°C / 0.2°C ; 0.8 °C/ 1.6 °C/ 0.8 °C
+ Acceptable excitation in 0.1 - 3.0 mA
- Pt500 385
+ Range:  200 to + 500 °C / +500 to + 630 °C
+ Accuracy 4-wire input/ 2-wire/3-wire input/ Output: 0.3°C / 0.4°C / 0.6°C ; 0.9°C / 0.3°C / 0.4 °C
+ Acceptable excitation in 0.05 - 3.0 mA
- Pt1000 385
+ Range:  - 200 to + 100 °C / +100 to + 630 °C 
+ Accuracy 4-wire input/ 2-wire/3-wire input/ Output: 0.2°C / 0.2°C / 0.4°C; 0.5°C / 0.2 °C/ 0.2 °C
+ Acceptable excitation in 0.1 - 3.0 mA
- Pt100 JIS
+ Range: + 200 to + 630 °C
+ Accuracy 4-wire input/ 2-wire/3-wire input/ Output: 0.2°C / 0.5 °C/ 0.3 °C
+ Acceptable excitation in 0.1 - 3.0 mA
- General specification
+ Power supply : 6 x 1.5 V
+ Dimension : 205 x 97 x 45 mm/ 472 g
+ Mains power supply: Input: 100 V – 240 VAC, 50–60 Hz 1.8 A ; Output: 12 VDC, 2 A MAX
 
Delivered with 2 pairs of test wires and crocodile clip, 6 AAA 1.5 V batteries, 1 operating manual in 5 languages and 1 carrying bag
Accessory: Mains power supply P01103057

datasheet
user guider

Calibrators offer two major functions: simulation and measurement.
SIMULATION: The first function involves simulating a signal or generating a quantity with precisely-defined values. These quantities are applied upstream of the measurement chain.
MEASUREMENT: The second function, involving the measurement of the signal. If there isa panel meter at the end of the chain,you can compare its value with the one indicated by the calibrator. If there is no panel meter, the simulated input value is compared with the signal measured by the calibrator
-The whole range of calibrators
(C.A 1621, C.A 1623 and C.A 1631) is designed to accurately measure, generate or simulate all the low-level signals encountered in industry.
Because they are simple to use, they are ideal for on-site work. Due to the wide variety of sensors recognized, these  instruments are suitable for all your applications.
Powered by batteries, these calibrators are stand-alone instruments offering a long battery life. When connected to the mains, they can be used continuously for as long as necessary!

获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家