Criteria for selecting hydrolysis particle counter

2025年03月22日 08時36分49秒

How to choose the right particle counter for your clean room? Expensive equipment is not always the best, but it is important that it meets your needs. If you often have to monitor the environment in a factory that produces electronic components, chips or mobile phones, there will certainly be specific criteria to consider. Output connection ports, air flow, particle size to be measured, all of these factors will determine which device is suitable for your system. So what are the most important features to choose? The following article will help you understand better!

What is a particle counter?

Each type of particle counter has a specific volumetric flow rate, which directly affects the airflow velocity passing through the device. Typically, this flow rate is measured in cubic feet per minute (CFM) or liters per minute (LPM). During calibration, each counter is set to a specific flow rate, and the accuracy of particle size measurement depends on this parameter.

To meet cleanroom classification standards (ISO 14644), a particle counter collects an air sample based on volume to provide statistical data.

Standards for selecting clean room particle counters

If compliance with ISO 14644-1:2015 cleanroom standards is required, the flow rate of the particle counter will determine the sampling duration. For example, a device with a flow rate of 1 CFM needs 35 minutes to sample 1 cubic meter of air. However, a counter with a flow rate of 50 LPM reduces this time to just 20 minutes.

In cleanroom monitoring, each ISO classification defines particle concentration limits. For an ISO Class 5 cleanroom measuring 0.1 µm particles, the particle counter must be capable of detecting concentrations of at least 100,000 particles/m³ (equivalent to 2,841 particles/ft³). If monitoring the same class with a particle counter measuring 0.3 µm particles, the requirement increases to over 102,000 particles/m³ (290 particles/ft³).

One crucial point to note: the ISO standard does not specify a 0.1 µm particle concentration requirement for ISO Class 6 or higher. Therefore, using a particle counter for particles smaller than 0.1 µm is unnecessary in these cases.

Reference:

Particle Counter CEM DT-9851M

Particle Counter CEM DT-9881M

See more: EMIN Hydrolysis Particle Counter

Cleanroom Monitoring: Which Method Is Right for You?

Regular clean room monitoring

Periodic cleanroom monitoring is a mandatory requirement. This process occurs at predetermined intervals, not exceeding 60 minutes during system operation. One of the most cost-effective solutions is utilizing a Manifold system, especially if installed during the cleanroom construction phase.

A portable particle counter can be easily connected to the Manifold system at any time. This system consists of 16 or 32 sampling ports, collecting data from multiple locations and channeling it into a single particle counter in a sequential process. However, due to its cyclic sampling mechanism, there is a risk of missing sudden airborne fluctuations that the system may not capture in real time.

Continuous clean room monitoring

Continuous (online) monitoring ensures uninterrupted data collection, allowing for timely detection of any incidents. Sampling intervals can be adjusted flexibly, but higher frequency provides better accuracy.

Short sampling intervals generate large volumes of data, which can overload the system. Therefore, standard sampling intervals typically range from 1 to 10 minutes. The decision between continuous or periodic monitoring depends on budget constraints and factory infrastructure.

For detailed tracking of particle movement, the ideal approach is to assign a dedicated particle counter to each monitoring point, though this can be costly. Alternatively, if continuous sampling is not required but sufficient data is needed to assess air quality trends, a Manifold system offers a cost-effective solution. However, this method is not suitable for particles larger than 5 µm due to data transmission limitations. nếu là anh, thìa huownshg 

Connection interface of the particle particle counter

When choosing a particle counter, you should also consider the ability to connect to a central monitoring system. If the cleanroom has an Ethernet port (10 base-T or 100 base-T), prioritize a machine with network connectivity for easy integration. If the system uses serial communication, you should choose a machine that supports RS-232 or RS-485 communication standards to ensure stable data transmission.

Particle testing in laboratory environment

Laboratory particle count testing applications typically do not require compliance with ISO cleanroom standards. Instead, they focus on quantifying particles within a specific size range to determine pass or fail criteria based on the test requirements.

Selecting an appropriate particle counter depends on the type of material or sample being analyzed. Laboratories must define the critical particle size threshold (measured in µm) and the maximum allowable concentration.

High-flow-rate particle counters are often preferred due to their ability to reduce sampling time and enhance data accuracy. Since most laboratory tests target ultrafine contaminants below the micrometer scale, the range of suitable instruments is narrowed, simplifying the selection process.

相关新闻

Why Is Associated Research Considered an Icon of Hipot Technology?
2026年07月14日 09時05分30秒

When it comes to electrical safety testing equipment specifically Hipot testers—many QA/QC and testing engineers are familiar with the globally renowned brand Associated Research. A subsidiary of the parent company IKONIX, Associated Research (ARI) invented the world's first commercial Hipot tester.

What Is the Standard 3-Phase Voltage? How to Read Cabinet-Mounted Meters Without Errors
2026年07月13日 15時09分18秒

Voltmeters are almost invariably found on the panels of electrical cabinets in factories, power stations, or generator systems. A glance lasting just a few seconds allows an operator to determine whether the power supply is stable or showing signs of abnormality. However, not everyone understands the significance of the displayed readings, particularly regarding three-phase electrical systems.

Distinguishing between ARI’s Hypot, HypotUltra, and HypotMax: Which device is right for your factory?
2026年07月13日 09時48分30秒

Few people realize that the concept of the "Hipot Test"—or electrical safety testing—has long existed within the electrical equipment manufacturing industry. Driven by the growing need to verify the voltage withstand capabilities of conductors, motors, and electrical devices, the industry required a robust and virtually fail-safe method to detect insulation weaknesses before products reached the market.

What Is Resonant Frequency? Why Can a Single Frequency Range Damage the Entire Product?
2026年07月10日 15時10分46秒

A device may operate stably for months, yet suffer cracks, broken solder joints, or snapped component leads after being transported. What surprises many engineers is that the vibration amplitude during transport is no greater than usual. The root cause lies in a specific frequency range where the product's structure absorbs energy most intensely, triggering resonance and causing continuous stress concentration at a single point.

获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家