Digital MultiMeter Calibration Procedure

2021年04月06日 10時26分04秒

**1.Range of application** \- DC Voltage 1000V \- AC Voltage 1000V, Frequency 100kHz \- DC Current 30A \- AC Current 30A, Frequency 10kHz \- Resistor 100

1.Range of application

- DC Voltage 1000V

- AC Voltage 1000V, Frequency 100kHz

- DC Current 30A

- AC Current 30A, Frequency 10kHz

- Resistor 1000MΩ

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s): Multifunction Calibrator (MFC)

Meatest M142 Multifunction Calibrator 15 ppm

- DC Voltage: 0µV - 1000V.

- AC Voltage: 0,1mV - 1000V, Frequency: 20Hz - 100kHz.

- DC Current: 0µA - 30A.

- AC Current: 1µA - 30A, Frequency: 20Hz - 10kHz.

- Resistor: 0 Ω - 1GΩ.

3.2 Auxiliary Instrument(s)

3.2.1 Testo Hygro-Thermometer 608-H1

Testo 608-H1 - Thermal Hygrometer (Part Number 0560 6081) | TEquipment

- Range of Measurement: (0 ÷ 50) °C; Accuracy: ±0,5 °C

- Range of Measurement: (10 ÷ 95) %RH; Accuracy: ±3 %RH

3.2.2 Dedicated Measuring Probe

- Probes with small electromotive force.

4. Conditions for Calibration

- Temperature: (23 ± 5) °C.

- Humidity: no higher than 80% RH.

5. Prepare for Calibration

- Make sure the power supply to the DMM meets the requirements specified in the catalog, the fuses, the protective circuit are still working properly.

- The DMM should be placed in the calibration environment for at least 8 hours.

- MFC and DMM must be turned on and dried for at least 30 minutes prior to calibration.

6. Begin the calibration

a. Outside Check

b. Technical Check

c. Measurement Check

    1. General requirements

    2. Begin the DMM Calibration

        a. DC Voltage

        b. AC Voltage

        c. DC Current

        d. AC Current

        e. Resistor

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated DMM is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of DMM is 12 months.


For calibration of DMM, click here.

For further information about Digital MCalibration service, please contact

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220


相关新闻

How do HRC, HRB, and HRA scales differ in hardness testing?
2026年06月15日 14時30分32秒

HRC, HRB, and HRA all belong to the Rockwell hardness system but are applied to different material groups. Understanding the differences between HRC, HRB, and HRA helps in selecting the correct testing method, accurately reading specifications, and avoiding unnecessary errors in material quality assessment.

Common Errors in Fiber Loss Optic
2026年06月13日 09時24分06秒

Optical power meter used together optical light source, can be applied to determine optical fiber link attenuation — a parameter that reflects the overall quality of the fiber optic cabling system

获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家