直接联系认证专家

Display: Easy to Read, Yellow Text on Black Background
Suggested Operating Ranges:
Standard High Range: UVA, UVB, UVV - 100mW/cm² to 10W/cm² / UVC - 10mW/cm² to 1W/cm²
Mid-Range: UVA, UVB, UVV -10mW/cm2 to 1W/cm2 / UVC: 1mW/cm2 to100mW /cm2
Low Power: UVA, UVB, UVV - 1mW/cm² to 100mW/cm² / UVC - 1mW/cm2 to 100mW/cm2
The suggested Operating Ranges are where the instrument performs best. Units will “turn on” and display data at irradiance values much lower than the suggested Operating Ranges.
Accuracy +/- 10%; +/- 5% typical plus ±0.2% of full scale Typical +/- 5% or better
Spectral Ranges: Four channel monitoring of UVA (320-390 nm), UVB (280-320nm) , UVC (250-260nm) and UVV (395- 445nm)
Spatial Response: Approximately cosine, “Lambertian”
Operating Temperature: 0-75°C Internal temperature; tolerates high external temperatures for short periods (audible alarm indicates when temperature has exceeded tolerance)
Smooth Modes:
Smooth ON: Effective Sample rate of 25 samples/second
Smooth PROFILER: Effective Sample rate of 128 samples/second
Smooth OFF: Effective Sample rate of 2048 samples/second
Time-Out Period: 2 minutes DISPLAY mode (no key activity)
Battery/Battery Life: Two user-replaceable AAA Alkaline Cells/Approximately 20 hours with the display “on”
Dimensions: 4.60 x 0.50 inches; 117 mm x 12.7 mm (D x H)
Weight: 10.1 ounces (289 grams)
Instrument Materials: Aluminum, stainless steel
Carrying Case Material/Weight: Cut polyurethane interior, scuff resistant nylon exterior cover/9 ounces (260 grams)
Carrying Case Dimensions: 10.75 x 3.5 x 7.75 inches; 274 x 89 x 197 mm (W x H x D)