ETL Prüftechnik 211032 HVE Module – Measurement Range Change (0.001 µA – 1000 mA)
Modification of hardware and software to achieve the following measuring ranges:
Current high voltage DC:
< 1 µA / 0.001 µA / ±1% vMW ±20 digits
10 µA / 0.001 µA / ±1% vMW ±3 digits
100 µA / 0.01 µA / ±1% vMW ±3 digits
1 mA / 0.0001 mA / ±1% vMW ±3 digits
5 mA / 0.001 mA / ±1% vMW ±3 digits
1 A / 0.0001 A / ±1% vMW ±3 digits
Voltage High voltage DC:
1000 V / 0.1 V / ±0.5% ±3 digits
8000 V / 1 V / ±0.5% ±3 digits
Current high voltage AC, RMS measurement:
< 1 µA / 0.001 µA / ±1% vMW ±20 digits
10 µA / 0.001 µA / ±1% vMW ±3 digits
100 µA / 0.01 µA / ±1% vMW ±3 digits
1 mA / 0.0001 mA / ±1% vMW ±3 digits
5 mA / 0.001 mA / ±1% vMW ±3 digits
1 A / 0.0001 A / ±1% vMW ±3 digits
Voltage High voltage AC:
1000 V / 0.1 V / ±0.5% ±3 digits
6000 V / 1 V / ±0.5% ±3 digits
Legend:
Measuring range / resolution (digit) / accuracy in % of measured value ± counts of least significant digit)