(1) Operation display:5.7 inch lattice localization color LCD
(2) Temperature control region:6 Road
(3) Temperature range: room temperature 8 ℃ ~ 400 ℃,Incremental: 1 ℃, Accuracy: ±0.1℃(4) Temperature programmed order number: 16 order
(5) Gas Control: Mechanical valve control mode, and electronic pressure flow control modefor option.(6) External event: 4 channels;
(7) Injector type: Packed column sample, capillary tube, six-way valve gas sampling, automaticheadspace sampling for option.
(8) Detector: FID, TCD, ECD(optional), FPD(optional)
(9) Boot into the sample: manual, automatic optional
(10) Communication interface: Ethernet: IEEE802.3
FID(flame ionization detector)
Detection limit: ≤2×10-11g/s (Hexadecane/isooctane)
Baseline noise: ≤5×10-14A
Baseline shift: ≤1×10-13A/30min
Linear range: ≥ 106
TCD(thermal conductivity detector)
Sensitivity: S≥2500mV•ml/mg (benzene/methylbenzene) (enlarge 1,2,4,8 times for option)
Baseline noise: ≤20μV
Baseline shift: ≤30μV/30min
Linear range: ≥104
ECD (electron capture detector) optional
Detection limit: 1×10-13g/ml (Hexachlorocyclohexane/isooctane)
Baseline noise: ≤0.03mV
Baseline shift: ≤0.2mV/30min
Linear range: 103
Radioactive souce: 63Ni
FPD (flame photometric detector) optional
Detection limit: (S)≤5×10-11g/s, (P)≤1×10-12g/s(parathion-methyl/ absolute ethyl alcohol)
Baseline noise: ≤3×10-13A
Baseline shift: ≤2×10-12A/30min
Linear range: Sulphur≥102, phosphorus≥103