For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Gastec 174L 快速1,3-丁二烯检测管 CH2:CHCH:CH2(2.5~100ppm)

联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Indicator: 10 to 100 ppm
Measuring Range: 
+ 2.5 to 5 ppm, Number of Pump Strokes 8
+ 5 to 100 ppm, Number of Pump Strokes 4
Sampling Time: 3 minute per pump stroke
Colour Change: Pale yellow → White
Reaction Principle: CH2:CHCH:CH2 + (NH4)2MoO4 + PdSO4 → White product 
Coefficient of Variation: 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Performance
Measuring Range
2.5 to 5 ppm 5 to 100 ppm
Number of Pump Strokes
8 4
Correction Factor
1/2 1
Sampling Time 3 minutes per pump stroke
Detecting Limit 0.5 ppm (n=8)
Colour Change Pale yellow → White
Reaction Principle CH2:CHCH:CH2 + (NH4)2MoO4 + PdSO4 → White product
Coefficient of Variation 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Acetylene, Hydrogen cyanide   + White
Carbon monoxide, Hydrogen   + Blue (whole layer)
Hydrogen chloride   + Pink
Hydrogen sulphide   + Black
Ethylene, Propylene, Butylene   + Blue
Butane, Pentane   + Blue (whole layer)
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
1,3-Pentadiene 8.5 4 42.5 to 850 ppm
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
2 ppm(2014) - 2.0 to 12%
获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家