Indicator: 2~50%
Measuring Range:
+ 1 ~ 2%, 1 Number of Pump Strokes
+ 2 ~ 50%, 1/2 Number of Pump Strokes
Sampling Time: 1 minute per 1/2 pump stroke
Colour Change: White → Blackish brown
Reaction Principle: 5CO + I2O5 + H2S2O7 → I2
Coefficient of Variation: 10% (for 2 to 15 %), 5% (for 15 to 50 %)
Nr. of tests: 10
Pump: GV-100S/110S
Catalog
Manual
| t of Variation |
10% (for 2 to 15 %), 5% (for 15 to 50 %) |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Acetylene |
0.2% |
Puls error |
Blackish brown |
Olefines (Ethylene, Propylene) |
2/3 |
Puls error |
Brown |
| Butane |
1% |
Puls error |
Pale brown (whole layer) |
| Propane |
20% |
Puls error |
| Calibration gas generation |
High pressure gas cylinder method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 25 ppm(2014) |
- |
12.5 to 74% |