Indicator:
Measuring Range:
+ 1 to10 ppm, Number of Pump Strokes 2 to 10
+ 10 to 120 ppm, Number of Pump Strokes 1
+ 120 to 240 ppm, Number of Pump Strokes 1/2
Sampling Time: 45 seconds per pump stroke
Colour Change: White → Brown
Reaction Principle: H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation: 5% (for 10 to 120 ppm)
Nr. of tests: 10
Pump: GV-100S/110S
Catalog
Manual
Manual tube
Performance
| Measuring Range |
| 1 to10 ppm |
10 to 120 ppm |
120 to 240 ppm |
|
| Number of Pump Strokes |
|
| Correction Factor |
|
| Sampling Time |
45 seconds per pump stroke |
| Detecting Limit |
0.2 ppm (n=10) |
| Colour Change |
White → Brown |
| Reaction Principle |
H2S + Pb(CH3COO)2 → PbS + 2CH3COOH |
| Coefficient of Variation |
5% (for 10 to 120 ppm) |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Nitrogen dioxide |
1/4 time |
 |
No |
| Sulphur dioxide |
1/1 |
+ |
No |
| Mercaptans |
|
No |
No |
| Hydrogen chloride |
|
Bleaching |
No |
| Calibration gas generation |
Permeation tube method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 1 ppm |
5 ppm |
4.0 to 44% |