Gastec 6AH 蒸汽探测管 H2O

Model6AH
原产地Japan
联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Indicator:  500 ~ 5000 ppm
Measuring Range:  500 ~ 5000 ppm
Sampling Rate: 300mL/min(300mL)
Sampling Time:  1 minute
Colour Change: Green → Purple 
Reaction Principle: H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O
Coefficient of Variation: 10% (for 500 to 1000 ppm), 5% (for 1000 to 5000 ppm)
Nr. of tests: 10

Catalog

 

Performance
Measuring Range 500 to 5000 ppm
Sampling Rate 300mL/min(300mL)
Correction Factor 1
Sampling Time 1 minutes
Detecting Limit 300ppm
Colour Change Green → Purple
Reaction Principle H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O
Coefficient of Variation 10% (for 500 to 1000 ppm), 5% (for 1000 to 5000 ppm)
Shelf Life 2 Years
Corrections for temperature Unnecessary for 0 to 40℃
Store the tubes at cool and dark place.
Calibration gas generation Bubbling method

Airtec tube - Compressed breathing air detector tubes

Airtec Tube for Cylinder / Compressed Breathing Air / Air Line Test.
Airtec tubes are a convenient and simple system for testing of a gas cylinder, compressed breathing air and air line. Industrial operations often produce, or are performed in the presence of harmful airborne contaminants. When self contained breathing apparatus or other devices are used for respiratory protection, the quality of the breathing air requires special attention. Contaminants entering the compressor or contaminants generated by the compressor or cylinder can be harmful to the worker and the respiratory equipment.
 




Measurement procedure:


1.Install a reducer pressure with gauge and flow metre to a cylinder, compressor or air line and adjust the flow metre the required setting.
  2.Break tips off a fresh detector tube in the tube tip breaker and insert a tube into a tube holder.
 
3.Attach the rubber tube holder to the flow metre outlet. Make certain the tube arrowon the tube pointing in a down direction.
  4.Turn on the cylinder or compressor and confirm the flow metre according to each Airtec tube specifications.
 
5.Time the flow rate with a stopwatch.
  6.As soon as sampling time has elapsed turn off the cylinder or compressor, and remove the tube from the tube holder and then read the colour change layer immediately.

 