GENERAL SPECIFICATIONS

OPERATING AMBIENT CONDITIONS

TEMPERATURE :18 °C to 28 °C

HUMIDITY: Up to 80 % RH

STORAGE CONDITIONS

TEMPERATURE: -20 °C to 50 °C

HUMIDITY: 0 % to 95 % RH (non-condensing)

ALTITUDE: 2000 m

CONTROLLED RANGE

TEMPERATURE : 5 °C to 55 °C

HUMIDITY@ 18°C 5 % to 95 % RH HUMIDITY@ 35°C 5 % to 80 % RH

HUMIDITY@ 23°C 5 % to 95 % RH HUMIDITY@ 40°C 5 % to 75 % RH

HUMIDITY@ 30°C 5 % to 85 % RH HUMIDITY@ 50°C 5 % to 60 % RH

CALIBRATION SYSTEM ACCURACY

TEMPERATURE:  ± 0.10 °C Or Better Typically ± 0.05 °C

HUMIDITY:  ± 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 %

CONTROLLER TYPE:  PID Controller

UNIFORMITY

TEMPERATURE : 0.10 °C Or Better Typically ± 0.05 °C

HUMIDITY : 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %

STABILITY

TEMPERATURE:  0.05 °C

HUMIDITY : 0.15 % RH @ 18 °C to 28 °C

RAMP / SOAK RATE OF CHANGE

Depends on ambient conditions

Temperature

High → Low 1.50 °C / minute (Typical)

Low → High 5.00 °C / minute (Typical)

Humidity

High → Low 5.00 % RH / minute (Typical)

Low → High 10.00 % RH / minute (Typical)

WARM-UP PERIOD

Ambient Condition : 30 Minutes Maximum

Cold : 30 Minutes

RESOLUTION

DISPLAY

Temperature 0.01 °C

Humidity 0.01 % RH

Dew Point 0.01 °C (Calculated)

USB

Temperature 0.01 °C

Humidity 0.01 % RH