直接联系认证专家
GENERAL SPECIFICATIONS
CALIBRATION
TRACEABILITY 17025 Traceable to NIST
OPERATING AMBIENT CONDITIONS
TEMPERATURE : 18 °C to 28 °C
HUMIDITY: Up to 80 % RH
STORAGE CONDITIONS
TEMPERATURE: -20 °C to 50 °C
HUMIDITY: 0 % to 95 % RH (non-condensing)
CONTROLLED RANGE
TEMPERATURE: 10 °C to 50 °C @ Ambient Temperature 23 °C
HUMIDITY@ 18°C 5 % to 75% RH HUMIDITY@ 35°C 5 % to 75 % RH
HUMIDITY@ 23°C 5 % to 85 % RH HUMIDITY@ 40°C 5 % to 70 % RH
HUMIDITY@ 30°C 5 % to 80 % RH HUMIDITY@ 50°C 5 % to 65 % RH
CALIBRATION SYSTEM ACCURACY
TEMPERATURE: ± 0.10 °C
HUMIDITY: ± 1.00 % RH
CONTROLLER TYPE : PID Controller
UNIFORMITY
TEMPERATURE: 0.20 °C
HUMIDITY : 0.40 % RH @ 23 °C
STABILITY
TEMPERATURE: 0.10 °C
HUMIDITY : 0.15 % RH @ 23 °C
RAMP / SOAK RATE OF CHANGE
Depends on ambient conditions and desiccant conditions
Temperature
High → Low 1.00 °C / minute (Typical)
Low → High 1.00 °C / minute (Typical)
Humidity
High → Low 3.00 % RH / minute (Typical)
Low → High 3.00 % RH / minute (Typical)
WARM-UP PERIOD
Ambient Condition : 30 Minutes Maximum
Cold : 45 to 60 Minutes
RESOLUTION DISPLAY
Temperature : 0.1 °C
Humidity : 0.1 % RH