直接联系认证专家
Be designed for dielectric testing of components and systems as well as testing of the electromagnetic compatibility of electronic systems and devices acc. to CCITT-K17/K20/K22, ITU-T/K44, IEC 61000-4-5 etc.
Mainframe
Microprocessor controlled touch panel: 7”, capacitive
Optical Ethernet Interface for remote control of the generator: optional
Interface for saving reports: USB
External Trigger input / output: Switch / 10 V
Connector for external safety interlock loop and external red and green warning lamps acc. to VDE 0104: 24 V=, 40mA
Mains power: 90V - 264V, 50/60 Hz
Dimensions: desk top case W * H * D: 450 * 180 * 500
Weight: 22 kg
High- Voltage Pulse Generator: acc. to IEC 61000-4-4: 2012
Impulse output voltage, adjustable ± 10%: 0.2 - 5 kV, 0.5 – 10 kV
Waveform of impulse output voltage: Selectable (acc. to IEC 60600) Tol. ± 30/±20%
Surge waveform, acc. IEC60600-1: 1.2/50µs, 10/700µs
Waveform of impulse output current at 10/700µs network, front time/tail time: 5/320 µs
Output polarity, selectable: pos / neg /alt
Trigger:
a) manual: push button
b) external: Trigger input switch
c) internal: automatic
Repetition time, selectable: 5-1000 s, 20-1000 s
Impulse voltage divider, built-in: v= 1000:1 ± 2 %, 50 W
Current Sense: acc. to IEC 61000-4-4: 2012
Threshold value, selectable: 1-3000 µAs(1,2/50µs)
Current sense working range: 0.2 kV - 5kV
Impulse voltage divider, built-in: ratio 1000:1 ± 5%
HV output: HV-OUT HV connector
Mains synchronous triggering, phase shifting: 0 - 360 °, step 1°
Accessories
Mains cable, key, operation instructions: