直接联系认证专家
pH
Range: 0.00 ~ 14.00pH
Resolution: 0.01pH
Accuracy: ±(0.05pH + 1d)
ORP
Range: -1800mV ~ +1800mV
Resolution: 1mV
Accuracy: ±(10mV + 1d)
Temperature
Range: -10.0°C ~ 120.0°C
Resolution: 0.1°C
Accuracy: ±(1°C + 1d)
Measurement of Water Sample Temperature: 0°C ~ 60°C
Isolated Output Signal: (4~20)mA DC, maximum load 750Ω
Output Upper and Lower Limits: pH: (0.00 ~ 14.00)pH optional, ORP: (-1800 ~ +1800)mV optional
Sample Flow Rate: (100 ~ 120)mL/min
Inlet Pressure: (7 ~ 140)kPa
Combination Electrode
Liquid Junction: PTFE double salt bridge
Reference Structure: Ag/AgCl
Temperature Compensation Unit: PT1000
Housing Material: PP
Outline Dimension: Ø30 × 170mm
Output Signal: Analog signal
Connector: Fork blade
Installation Mode: 3/4NPT
Pressure Range: ≤0.6MPa
Applicable Samples: Routine or contaminated samples
Operating Environment: Temperature: (-5 ~ 50)°C, Relative Humidity: ≤90%
Power Supply: AC 220V, 50Hz
Hole-opening Size: 140 × 140mm
Dimension (W × D × H): 166 × 166 × 110mm
Weight: 2kg