Parameter PFP7
Range 0 to 199.9ppm
Limits of detection Na 0.2ppm
K 0.2ppm
Li 0.25ppm
Ca 15ppm
Ba 30ppm
Reproducibility <1% coefficient of variation for 20 consecutive samples using 10ppm Na set to read 50
Linearity < 2% error when 3ppm Na/K and 5ppm Li are set to read 100
Stability < 2% drift over 5min when continuously aspirating 10ppm sample set to 50.0. Specificity Interference from Na/K and Li equal in concentration to test element will be <0.5%
Recorder output Nominal 1.00V for a reading of 100.0
Electrical supply 90-125V or 190-250V @50/60Hz
Air supply Moisture and oil free, 6 litre/min @ 14psi
Fuel Propane, butane, natural gas or LPG
Size (w x d x h) 420 x 360 x 300mm
Weight 8kg

Ordering Information

Part Code Description
500 701 PFP7 industrial flame photometer supplied with Na, K, Ba, Ca and Li filters, connecting hoses and clips, compressor plug and drain trap (230V/50Hz)

 


Note: This product will only work with the correct type of compressor and gas regulator.