直接联系认证专家

Sampling flow: 100L/min ±2.0%
- Resolution: 0.1L/min
Rotation speed of aerosol generator: 28000r/min ±500r/min
Diameter of rotating plate on aerosol generator: 38mm
Highest position of X1, Y1: 1000mm
Sampling membrane: Diameter 25mm,Aperture 3μm
Disturbing cylinder: Diameter 63mm,Length 1100mm
Noise:<65dB(A)
Working temperature: 0-40℃
Power supply: AC220V±10%,50Hz
Overall size:(length 450×width 380×height 720)mm
Overall power: < 400W
Overall weight: About 23kg
Features
8-inch colored high definition touch screen, the contents are intuitive and operation is easy.
Equipped special aerosol sampler, could capture special size aerosol.
Equipped calibrated special KI aerosol generator, the rotation speed is stable.
Four path independent high precision sampling module, automatic flow control, don’t need to adjust pressure to stabilize flow.
Test results could be exported to USB drive or printed by Bluetooth printer.
High precision liquid supply module, guarantee a stable liquid flow.
The automatic aerosol control port adopts PID arithmetic, control by real time feedback generator rotation speed.
Operator protection, product protection and cross protection test three test modes could start after just press the button.
Standards
YY 0569-2011: Class II biological safety cabinets
JJF 1815-2020: Calibration specification for of class II biosafety cabinet
DB52T 1254-2017: Technical practice for testing biosafety cabinets