直接联系认证专家
Performance Data Temperature
Temperature range [°C] :0…70
Temperature variation at 40 °C [± K]: 0.2
Temperature fluctuation [± K]: 0.1
Max. heat compensation at 40 °C [W]: 600
Performance Data Climate
Temperature range [°C]: 10…70
Temperature variation at 25 °C and 60 % RH [± K]: 0.2
Temperature variation at 40 °C and 75 % RH [± K]: 0.2
Temperature fluctuation at 25 °C and 60 % RH [± K]: 0.1
Temperature fluctuation at 40 °C and 75 % RH [± K]: 0.1
Humidity range [% RH] 10…80
Humidity fluctuation at 25 °C and 60 % RH: 1.5 ± % r.F.
Humidity fluctuation at 40 °C and 75 % RH: 1.5 ± % r.F.
Electrical data
Rated Voltage [V]: 200…230
Power frequency [Hz]: 50/60
Nominal power [kW]: 3.1
Unit fuse [A]: 16
Phase (Nominal voltage) [ph]: 1~
Outer dimensions
Width net [mm]: 1250
Height net [mm]: 1925
Depth net [mm]: 890
Wall clearance back [mm]: 100
Wall clearance sidewise [mm]: 300
Doors
Inner doors: 2
Unit doors: 2
Internal Dimensions
Width [mm]: 973
Height [mm]: 1250
Depth [mm]: 576
Further Dimension
Interior volume [L]: 700
Net weight of the unit (empty) [kg]: 312
permitted load [kg]: 150
Load per rack [kg]: 45
Environment-specific data
Energy consumption at 40 °C and 75 % RH [Wh/h]: 620
Sound-pressure level [dB(A)]: 53
Fixtures
Number of shelves (std./max.): 2/15