For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

KEYSIGHT E4982A LCR Meter (1 MHz ~ 3 GHz)

联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Frequency

Range

 1 MHz to 300 MHz (Option 030)

1 MHz to 500 MHz (Option 050)

1 MHz to 1 GHz (Option 100)

1 MHz to 3 GHz (Option 300)

Resolutionz: 1 kHz 1

Oscillator level

Cable length = 1 m

Power range (When 50 Ω LOAD isconnected to test port)

–40 dBm to 1 dBm

Current range (When SHORT is connected to test port)

0.0894 mArms to 10 mArms

Voltage range (When OPEN is connected to test port)

4.47 mVrms to 502 mVrms

Uncertainty (When 50 Ω LOAD is connected to test port)

(23 °C ± 5 °C)

± 2 dB (frequency ≤ 1 GHz)

± 3 dB (frequency > 1 GHz)

(5 °C to 40 °C)

± 4 dB (frequency ≤ 1 GHz)

± 5 dB (frequency > 1 GHz)

Resolution 0.1 dB (When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.)

Cable length = 2 m (When option 002 is used)

Power range Subtract the following attenuation from the power

Dimensions, weight: 

Weight Main unit: 13 kg, test head: 250 g with plate