Channel: 1
AC Hipot
Output Voltage: 10V~5000V
Voltage Resolution: 1V
Voltage Accuracy: ±(2% of setting +5V), no load
Output Frequency: 50/60Hz
Rated Current (Max.): 30mA
Leakage Current: 0~31mA
Lower Leakage Current Resolution: 0.001mA
Lower Leakage Current Accuracy: ±(1% of reading +5uA)
Measuring Time: 0.1~999s, 0 (Continuous)
Measuring Time Resolution: 0.1s
Ramp, Dwell, Fall Time: 0.1~99.9 sec
Arcing Detection: Level 0 for Off / Adjustable levels 1~9 (Level 1 is the most sensitive)
DC Hipot
Output Voltage: 10V~6000V
Voltage Resolution: 1V
Voltage Accuracy: ±(2% of setting +5V)
Rated Current (Max.): 10mA
Leakage Current: 0~11mA
Lower Leakage Current Resolution: 0.001mA
Lower Leakage Current Accuracy: ±(1% of reading +5uA)
Measuring Time: 0.1~999s, 0 (Continuous)
Measuring Time Resolution: 0.1s
Ramp, Dwell, Fall Time: 0.1~99.9 sec
Arcing Detection: Level 0 for Off / Adjustable levels 1~9 (Level 1 is the most sensitive)
Insulation Resistance
Output Voltage: 10V~6000V
Voltage Resolution: 1V
Voltage Accuracy: ±(2% of setting +5V)
Rated Current (Max.): 1mA
IR Upper Threshold: 200kΩ~100GΩ (Current measurement range 5nA ≤ I ≤ 1mA)
IR Resolution: 0.001MΩ
IR Accuracy:
50V: 200k~500MΩ, ±(5% of reading +30kΩ) / 500M~1GΩ, ±(10% of reading +20MΩ) / 1G~10GΩ, ±(25% of reading +100MΩ)
100V: 200k~200MΩ, ±(2% of reading +30kΩ) / 200M~1GΩ, ±(5% of reading +20MΩ) / 1G~20GΩ, ±(25% of reading +100MΩ)
500V: 500k~500MΩ, ±(2% of reading +30kΩ) / 500M~10GΩ, ±(5% of reading +20MΩ) / 10G~50GΩ, ±(20% of reading +100MΩ)
≥1000V: 1M~1GΩ, ±(2% of reading +30kΩ) / 1G~10GΩ, ±(5% of reading +20MΩ) / 10G~100GΩ, ±(20% of reading +100MΩ)
Measuring Time: 0.1~999s, 0 (Continuous)
Measuring Time Resolution: 0.1s
Ramp, Dwell, Fall Time: 0.1~99.9 sec
General
Remote Control: Test, Abort
Remote Output Signal: Pass, Fail, HV output, Testing
Safety Switch: When testing, you need to short-circuit the INTER LOCK on the rear of the instrument to output the test voltage
Built-in Storage: 64 sets
USB Host Storage: Storage setting files and update firmware; the number of saved data sets depends on the USB capacity
List Mode: Each group contains up to 8 steps
Interface: RS-232, Remote, USB Host, USB Device, EXT.I/O
Power Supply: 100Vac~240Vac / 50/60Hz ±5%
Power Consumption: 350VA
Display: 480 × 272, 4.3" TFT LCD
Environment: Temperature: 5℃~40℃, Humidity: 20~80%RH
Dimension (W×H×D): 340 × 88 × 369.6mm
Weight: 12Kg