Input

* Voltage (DC) (Range/ Input range/ Resolution/ Accuracy/ Remarks):

100 mV/ -10 to 110 mV/ 10 μV/ ±0.05% measured value ± 30 μV/ Input Resistance 1MΩ

* Thermocouple (Range/ Input range/ Resolution/ Accuracy/ Remarks):

+ R/ -40 to 1760°C/ 1°C/ ±0.05% measured value ±3°C (≤100°C); ±0.05% measured value ±2°C (>100°C)/ Input Resistance 1MΩ

+ S/ -20 to 1760°C/ 1°C/ ±0.05% measured value ±3°C (≤100°C); ±0.05% measured value ±2°C (>100°C)/ Input Resistance 1MΩ

+ B/ 400 to 1800°C/ 1°C/ ±0.05% measured value ±3°C (400 to 600°C); ±0.05% measured value ±2°C (>600°C)/ Input Resistance 1MΩ

+ E/ -200 to 1000°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ 

+ K/ -200 to 1370°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ

+ J/ -200 to 1200°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ

+ T/ -200 to 400°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ

+ N/ -200 to 1300°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ


Output

* Voltage (DC) (Range/ Output range/ Resolution/ Accuracy/ Remarks):

100 mV (Low)/ -10 to 110 mV/ 0.01 mV/ ±0.05% set value ±30 μV/ Max. output

current ±2 mA

1V (High)/ -0.10 to 1.10 V / 0.1 mV/ ±0.05% set value ±0.3 mV/ Max. output

current ±2 mA

* Thermocouple (Range/ Output range/ Resolution/ Accuracy/ Remarks):

+ R/ -40 to 1760°C/ 1°C/ ±0.05% set value ±3°C (≤100°C); ±0.05% set value ±2°C (>100°C)/ By using ITS-90 temperature scale 

+ S/ -20 to 1760°C/ 1°C/ ±0.05% set value ±3°C (≤100°C); ±0.05% set value ±2°C (>100°C)/ By using ITS-90 temperature scale 

+ B/ 400 to 1800°C/ 1°C/ ±0.05% set value ±3°C(400 to 600°C); ±0.05% set value ±2°C (>600°C)/ By using ITS-90 temperature scale 

+ E/ -200 to 1000°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale 

+ K/ -200 to 1370°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale 

+ J/ -200 to 1200°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale 

+ T/ -200 to 400°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale 

+ N/ -200 to 1300°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale


Standard delivery:

Thermocouple Calibrator – Model : Nagman 02

9V Alkaline Battery

Operation Manual

Test Leads

Traceable Calibration Certificate

Optionals:

Calibration Certificate from NABL Accredited –as per ISO/IEC 17025:2005 Laboratory 