直接联系认证专家
Measuring Range: 0…10, 0…100, 0…1000 ppmV
Accuracy:
10 ppm: +/- 0.5 ppm
100 ppm: +/- 1 ppm
1000 ppm: +/- 3 ppm @ 100 ppm; +/- 1 ppm @ 10 ppm
Output Resolution (4...20 mA): 1 ppmV
Lower Detection Limit (LDL): 1 ppmV
Sample Flow Rate (application dependent)
Flow-through / extractive: 100...500 ml/min (250 ml/min optimal) in a vented atmosphere
Direct insertion: Up to 6 m/s
Pressure Range: 900…1100 mBarabs
Response Time (T90): < 15 seconds @ 25 °C (77 °F) within selected range
Operating Temperature Range: +5°C…+45°C, (+31°F…+113°F)
Life Expectancy (application dependent): Up to 12 months
Humidity: 0…95 %rh non-condensing (with normal use)
Shelf Life: Up to 3 months
Calibration Interval (application dependent): 3...6 months
Electrical
Output Signal: 4…20 mA loop-powered
Electrical Interface: Industry standard M12
Power Supply: 24 V DC +/- 10 % (a suitable I.S barrier is required)
Mechanical
Ingress Protection: IP66 (NEMA4)
Housing Material: 316 Stainless steel
Process Connection: Flow-through (1/8” NPT), KF40 flange, Tri-clamp
Gas-wetted Materials: 316 Stainless steel
O-ring Material: Nitrile
Weight: 800g (28 oz)