For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

OMEGA CDH-287-KIT 电导率/电阻率/TDS/盐度计

ModelCDH-287-KIT
联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Conductivity 

Ranges:

0.00 to 19.99 µS/cm or 1.999 mS/cm

0.00 to 199.9 µS/cm or 19.99 mS/cm

0000 to 1999 µS/cm or 199.9 mS/cm

0.00 to 19.99 mS/cm or 1999 mS/m

00.0 to 199.9 mS/cm or 19.99 mS/m

000 to 1999 mS/cm or 199.9 mS/m

Accuracy: ±0.3% rdg

Temperature Compensation: 

Automatic, 0 to 50°C (32 to 212°F)

Total Dissolved Solids

Ranges:

 0 to 19.99 mg/L, 0 to 199.9 mg/L

 0 to 1999 mg/L, 0 to 1.999 g/L

 0 to 19.99 g/L, 0 to 132.0 g/L

 Accuracy: ±0.3% rdg

Resistivity

 Ranges:

 0 to 1.999 MΩ/cm or 0.019 MΩ/m 

 0 to 19.99 MΩ/cm or 0.199 MΩ/m

 Accuracy: ±0.3% of reading

Salinity

 Range: 2.0 to 42.0% salinity; automatic conversion from conductivity, using temperature relationship for seawater, in accordance with UNESCO, IASPO data

 Accuracy: ±0.3% rdg

 Temperature Compensation: Automatic, -2.0 to 35°C (28.4 to 95°F)

Concentration

 Range: 0 to 9999, automatic ranging, choice of units, background offset function

 Calibration: 4 point straight line interpolation

Temperature

 Range: -30.0 to 130.0°C (-22.0 to 266.0°F)

 Accuracy: ±0.3°C (0.5°F)

General Specifications 

Reference Temperature: 25°C (77°F),

selectable to 20°C (68°F)

Temperature Coefficient: Preset to

2%/°C; programmable from 0 to 5%/°C

Measurement Frequency: 3000 Hz

Recorder Output: ±200.0 mV

RS232

Display: 12.7 mm (0.5") LCD

Dimensions: 100 L x 180 W x 44 mm D (3.9 x 7.1 x 1.7")

Cable: 1.2 m (3.9')

Weight: 410 g (0.9 lb)

Power: 9V battery (included)

CDH-287-KIT is supplied with meter, glass dip-style conductivity probe with integral temperature sensor and cell constant of 1.0, sample bottle, 9V battery, calibration solution, rugged carrying case and operator’s manual.

Datasheet


User Manual


获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家