Width: 211 mm
Length: 480 mm
Accuracy: ± 0.5 °C
Polarity: Unipolar
RTD Type: 100/500/1000 Ω Pt sensor, 2-, 3- or 4-wire; α= 0.00385 or 0.00392
Input Type: Thermocouple, RTD, Thermistor, Current, Voltage
Resolution: 1°/0.1°; 10 µV process
Display Size: 102 mm (4 in)
Display Type: 4 Digit Single LED
Reading Rate: 3 samples/sec
Accuracy Note: ±0.5°C temperature; 0.03% rdg
Sampling Rate: 3 Hz (3 updates/second)
Step Response: 0.7 sec for 99.9%
A/D Conversion: Dual slope
Digital Filter: Programmable
Supply Voltage: 100 to 240 Vac
Impedance Input: 10 MΩ for 100 mV, 1 MΩ for 1 or 10 Vdc
Span Adjustment: 0.001 to 9999 counts, programmable
Number of Inputs: 1
Decimal Selection: 0, 0,001, 0,01, 0,1
Display Type Note: 4-digit, 7-segment LED 57.2 mm (2.25") or 101.6 mm (4.00") red, green and amber programmable colors for process variable, set point and temperature units
Enclosure Ratings: NEMA Type 1
Offset Adjustment: -1999 to +9999
Power Consumption: 22.5 W
Relative Humidity: 90% RH non-condensing
Supply Power Note: 100 to 240 Vac ±10%, 50/60 Hz 22.5 W
Supply Power Type: AC
Thermocouple Type: J, K, T, E, R, S, B, C, N, L
Input Configuration: Single-ended
Linearization Points: None
Set Point Adjustment: -1999 to 9999 counts
Common Mode Rejection: 120 dB
Normal Mode Rejection: 60 dB
Temperature Stability: RTD: 0.04°C/°C ;Thermocouple @ 25°C (77°F): 0.05°C/°C, cold junction compensation ;Process: 50 ppm/°C ;
Excitation Voltage Note: Optional in Place of Communication
Sensor/Transmitter Power: 24 Vdc @ 25 mA
Current Measurement Range: 0 to 20 mA
Voltage Measurement Range: 0 to 100 mV, 0 to 1 V, 0 to 10 Vdc
Warm-Up to Rated Accuracy: 60 min
Operating Temperature, Max: 40 °C
Operating Temperature, Min: 0 °C
Thermocouple Lead Resistance, Max: 100 Ω