直接联系认证专家

Size Range: 0.5 to 25.0 μm
Size Channels: Factory calibrated at 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm variable binning
Counting Efficiency: 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Flow rates: 1.0 CFM (28.3 LPM)
Concentration Limits: 1,500,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 3,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Light Source: Long life laser diode
Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction.
Count Modes: Real-Time Meter and graph, cumulative/differential count/m3 and count/ft3, and mass concentration (PM)
Count Alarms: 1 to 9,999,999 counts
Calibration: NIST traceable
Display: 4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Printer (Optional): External thermal printer available
Vacuum Requirements: External vacuum >15” (38.1 cm) of Hg
Airflow: Internally monitored
Number of Channels: 6
Custom Size Channels: Calibration for custom size channels available
Audible Alarm: Adjustable built-in alarm
Reports: ISO 14644-1, EU GMP Annex 1, FS 209E
Communication Modes: Ethernet, USB, Wireless 802.11 b/g (Optional), RS485 or RS232
Optional Environmental Sensor: Temperature (32-122°F, ±1°F / 0-50°C, ±0.5°C) and relative humidity probe (15–90% RH, ±2%)
Alarm: Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow rate
Standards: ISO 21501-4 and JIS B9921
Calibration: Recommended minimum once per year
External Surface: Stainless steel
Dimensions (L x W x H): 5.2 x 2.3 x 8.9 in (13.3 cm x 5.8 cm x 22.6 cm) includes barb fittings
Weight: 2.6 lb (1.18 kg)
Accessories: User manual + Instrument Management Software (IMS) on USB flash drive, isokinetic probe, temperature relative humidity sensor, zero-count filter, USB cable, and power supply
Optional Accessories: Printed manual, carrying case, external printer, barbed fittings, and tubing
Buffer Memory: Stores up to 45,000 sample records with particle counts, environmental data, locations, and timestamps
Sample Locations: Up to 1,000 locations 20 characters long
Sample Time: 1 second to 99 hours
Power: 110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions: 41-104ºF (5-40ºC) / 20% to 95% non-condensing
Storage Conditions: 32-122ºF (0-50ºC) / Up to 98% non-condensing