Impact Energy: 5.00±0.25J
Tolerance: ±5%
Equivalent mass: 1.7kg±2%
Trigger force: <10N
Total mass: -
IK08/5J spring hammer test apparatus is designed to meet the test requirements of IEC 62262 and IEC 60068-2-75. This is not a stanard recommended equipment, as spring hammer is not recommended for severities 2J above. But in practice, the 5J spring hammer has been improved to make it easy to operate and not harm for operator. The structure of 5J spring hammer is designed according to IEC60068-2-75 figure E.1, the imapct energy can pass the calibration of the third-party lab.