For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

PEGO PG-SYDY 残压测试仪 (0~250V/50Hz (AC), 10~199.9V)

ModelPG-SYDY
联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Output voltage: 0~250V/50Hz (AC)

Output current: ≤10A

Output power: ≤3KVA

Power supply time

1~99s (The equipment automatically judges that it is disconnected at the voltage peak for testing)

Test voltage: 10~199.9V( resolution: 0.1V, accuracy: ±(reading*1%+1V))

Test time: Test after disconnecting for 1 second

Number of tests: 1~99 times

Test impedance: 1000MΩ

Alarm range: Settable (normally 34V or 60V)

Judgement: Sound and light alarm, and cut off the voltage at the same time

Working power: 220V/50Hz

The residual voltage tester is designed according IEC60601-1:1988 clause 15 and IEC60335-1:2016 clause 22.5. The equipment has a high enough test impedance, it uses the method of simulating the disconnection of the plug to disconnect the appliance from the supply mains at the instant of voltage peak. It is used for the residual voltage test of electrical appliance like medical equipment and household appliances.

Working principle

The appliance is supplied at rated voltage, any switch is then placed in the off position and the appliance is disconnected from the supply mains at the instant of voltage peak, 1s after disconnection, the voltage between the pins of the plug is measured with an instrument that does not appreciably affect the value to be measured, the voltage shall not exceed 34V as per IEC60335-1/60V as per IEC60601-1.