PEGO RK2672CY 耐压测试仪(医疗设备) ((0.00-5.00)KV, 100mA)

ModelRK2672CY
联系我们
安全结算
品质保证
轻松更换与退货
支持配送

ACW

Output voltage range: (0.00-5.00)KV

Maximum output power: 500VA (5.0KV, 100mA)

Maximum rated current: 100mA

Current gear: 2mA, 20mA, 100mA

Output waveform: Sine-wave

Output waveform distortion: ≤5% (no-load or pure resistive load)

Test time: 0.0s-999s 0=continuous testing

DCW

Output voltage range: (0.00-5.00)KV

Maximum output power: 100VA (5.0KV, 20mA)

Maximum rated current: 20mA

Current gear: 2mA, 20mA

Ripple factor: ≤5%

Test time: 0.0s-999s 0=continuous testing

Voltmeter

Range: (0.00-5.00)KV

Accuracy: ±(5%+3 digits)

Resolution: 10V

Display value: RMS

Ammeter

Range

Range 1: 0.1mA~2mA, range 2: 2mA~20mA, range 3: 20mA~100mA

Accuracy: 2mA: 1μA, 20mA: 10mA; 100mA: 0.1mA

Resolution: ±(5%+3 digits)

Timer

Range: 0.0s-999s

Resolution: 0.1St

Accuracy: ±(1%+50ms)

The RK2672CY medical withstand voltage tester has all the functions of the conventional withstand voltage tester, and has added an arc (flashover) detection function. Through an external oscilloscope, it can detect the “flashover” phenomenon intuitively, accurately, quickly and reliably. When the electrical equipment under test has no "flashover" phenomenon, the oscilloscope will display a stable "Lissajous figure (that is, a closed circle)", if the electrical equipment under test has a "flashover" phenomenon, the edge of the Lissajous waveform will appear jagged, indicating that there is a "flashover" inside the electrical equipment under test.

This product is also suitable for all kinds of electrical equipment that need to detect "flashover" defects. It is an indispensable withstand voltage tester for medical electrical equipment manufacturers, maintenance departments, users, product quality inspection departments, and technical supervision departments.