直接联系认证专家

Interferometer: Cube-corner Michelson interferometer
Beam splitter: Multilayer Ge coated KBr
Detector: Room temperature DLATGS module (standard)
IR Source: High intensity, long lifetime, air-cooled IR source
Wavenumber Range: 7800cm-1~350cm-1
Resolution: 0.85 cm-1
Signal to noise ratio: Better than 20,000:1 (RMS value, at 2100cm-1 ~ 2000cm-1 or 2100cm-1 ~ 2200cm-1, resolution: 4cm-1, detector: DLATGS, 1 minute data collection)
Wavenumber Accuracy: ±0.01 cm-1
Scanning Speed: Microprocessor control, different scanning speed selectable.
Software: Main FTWin Suite software workstation, compatible
Communication:
Ethernet interface (standard)
WIFI wireless communication (optional)
Data Output: Standard data format, report generation
Status Diagnosis: Power on
Certification: FDA 21 CFR Part11 compliance,IQ/OQ/PQ(Available as an Accessory)
Environment Conditions: Temperature: 10℃~30℃,humidity: less than 70%
Power Supply: 220V±22V,50Hz±1Hz; 110v 50Hz±1Hz
Accessories: Comprehensive Range of
Dimensions & Weight: 520 x 435 x 255mm, 24kg