Measurement Range: 0–300,000 cells/mL
Detection Limit: 200 cells/mL
Measurement Accuracy: ±10% of the signal level corresponding to 1 ppb Rhodamine B Dye
Resolution: 20 cells/mL
Repeatability: ±3%
Pressure Range: ≤0.4 MPa
Calibration: Deviation value calibration, slope calibration
Requirements:
Suggest multipoint monitoring for uneven distribution of Blue-Green algae in water.
Water turbidity should be below 50 NTU.
Main Material:
Body: SUS316L + PVC
Cover: POM
Cable: PUR
Power Supply:
AC: 85–500 VAC
DC: 9–36 VDC
Output: 3-way 4–20 mA
Communication Protocol: MODBUS RS485
Storage Temperature: -15–50°C
Measuring Temperature: 0–45°C (non-freezing)
Dimension:
Sensor: Dia 30 mm × Length 223 mm
Transmitter: 145 × 125 × 162 mm (L × W × H)
Weight:
Sensor: 0.55 kg
Transmitter: 1.35 kg
Protective Rate:
Sensor: IP68 / NEMA6P
Transmitter: IP66 / NEMA4X
Cable Length: Standard 10 m, maximum may be extended to 100 m
Relay: Three-way relays, programmed response parameter and response value