Test device: Automatic range conversion, true RMS measurement
Medical GB9706.1-2020 Standard
Leakage current measurement range AC: I (3~99.9)uA Resolution: 0.1uA
Ⅱ (100.0~999.9)uA Resolution: 0.1uA
Ⅲ (1000~9999)uA Resolution:1uA
Patient current, patient auxiliary current: DC Measurement range: (3~999.9)uA Resolution: 0.1uA
Measurement accuracy: ±5%+3 digits Note: The accuracy range is current > 10uA or more
Frequency response range: DC ~ 1MHz Measurement impedance circuit (MD): in line with GB9706.1-2020
Laboratory GB4793.1-2007 Standard
Leakage current measurement range AC: I (3~99.9)uA Resolution: 0.1uA
Ⅱ (100.0~999.9)uA Resolution:0.1uA
Ⅲ (1000~9999)uA Resolution:1uA
Measurement accuracy: ±5%+3 digits Note: The accuracy range is current > 50uA or more
Measuring Impedance Circuit (MD): Compliant with GB4793.1-2007
The output voltage
Measure the voltage output range of the power supply: 0V ~ 300V, the resolution is 1V, the accuracy: ± (5% + 2 words)
Output voltage capacity: 3000VA
Current upper limit setting
Range: (3~9999)uA
Resolution: 1uA Accuracy: ±(4%+3 words)
The accuracy range is current>10uA
Timing device
Range: (3~9999)S
Resolution: 1S
Accuracy: ±5%
Operating temperature: 0-40℃ ≤75%RH
Power requirements: 220±10% 50Hz/60Hz±3Hz
interface: Standard RS232, PLC
Memory group: five sets of memory
Screen: LCD2004C
Form factor(D×H×W): 430mm×150mm×350mm
weight: 30KG
Instrument power consumption: <50W
Random standard accessories: power cable RK00001、RS232 Communication Cable RK00002、RS232 to USB cable RK00003、16G U disk (instruction manual)、Cable interface transfer driver CD、test leadRK00049/RK00050
Optional accessories: host computer