For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

RIGOL RP1004C Current Probe (DC ~ 100MHz)

联系我们
安全结算
品质保证
轻松更换与退货
支持配送

Bandwidth: DC to 100 MHz (-3 dB), refer to Appendix 1 Amplitude Characteristics (RP1004C)

Rise Time: ≤3.5 ns

Linear Response Range of Input Current: 30 Arms, refer to Appendix 2 Relation between Max Input Current and Frequency (RP1004C)

Maximum Peak Current: 50 A peak, non-continuous

Gain: 0.1 V/A

Gain Accuracy:

±1.0%rdg ±1 mV, ≤30 A

±2.0%rdg, 30 A to 50 A peak

(DC and 45 Hz to 66 Hz, within the linear response range of the input current)

Noise: ≤2.5 mArms (for 20 MHz band measuring instrument)

Input Impedance: Refer to Appendix 3 Input Impedance (Typical) (RP1004C)

Gain Accuracy Drift: ≤±2% (0 ℃ to 40 ℃ temperature, input of 50 Hz, 30 A)

Rated Power: 5.3 W

Supply Voltage: +12 V ±0.5 V

Maximum Rated Voltage: 300 V, CAT I (insulated conductor)

Gain Accuracy Period: 1 year (opening/closing up to 10,000 times)

Effect of External Magnetic Fields: ≤5 mA (DC and 60 Hz, magnetic field of 400 A/m)

Datasheet

获取优惠更新

获取专属批量折扣、批发价格更新和新产品通知,直接发送到您的邮箱。

订阅即表示您同意我们的服务条款隐私政策

快速支持

直接联系认证专家