Resistance thermometers (RTD):
Generation and measurement of RTD signals: Pt50, Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50, Ni100, Ni120, Ni1000
Ω-generator function : 0...4000 Ω
Ω-measurement function : 0...4000 Ω
Accuracy (of reading + const.) : ±0.012 %
Selectable temperature unit : °C / °F
Measurement of multi-wire connections : 2/3/4
Thermocouples (TC):
Generation and measurement of TC signals : J, K, T, R, S, B, N, E, U, L
mV-generator function: 0...100 mV
mV-measurement function : 0...100 mV
Accuracy (of reading + const.) : ±0.013 %
Selectable temperature unit : °C / °F
Internal comparison point : ±0.3 °C
Current (mA):
Loop current signal generation: 0(4)...24 mA
Loop current signal measurement : 0(4)...50 mA
Accuracy (of reading + const.): ±0.0175 %
Current loop supply : 24 V ±10 %, 22 mA
HART Communication protective resistor: 250 Ω
Voltage (V):
Voltage signal generation : 0...20 V
Voltage signal measurement : 0...50 V
Accuracy (of reading + const.) : ±0.015 %
Continuity (O/C):
Continuity measurement : O / C
Switching threshold „open“ : 1 kΩ
Frequency and pulse (Hz):
Frequency and pulse signal generation : 0.01 Hz...10 kHz
Frequency and pulse signal measurement : 0.01 Hz...20 kHz
Accuracy (of reading + const.) : ±0.005 %
Pressure signals (bar):
Pressure measurement with external pressure module: ✓
Editable pressure units : ✓
Connection via DIN socket : 5-pin
General data:
Features:
Sensor connection : 8 safety sockets (2 channels) incl. push lock, 4 sets with safety instrument leads (4 mm) and alligator clips (optional)
Power supply : Rechargeable battery including power adapter, 230 VAC / 12 VDC
Operator guidance : Menu with pull-down windows; Separate channels for parallel signal processing
Display : Multifunction display, double-LCD, 2 x 6 digit, illuminated
Housing:
Version : ABS plastic with protective cover
Dimensions : 110 x 210 x 50 mm (H x W x D)
Weight: Approx. 900 g