Measurement Range:
1 to 200 MΩ, for resistance
3 to 600 VAC RMS, for voltage
Test Voltage for Resistance: 500 VDC 10%
Accuracy* & Calibration:
Voltage readings within 5% of reading ± 1 count
Resistance readings within 3% of reading ± 1 count
Suggested Calibration Interval: 12 months
End-of-interval confidence factor: 72%
Recommended Calibration Cycle: 1 year
Controls and Display Single:
“TEST” switch (mode selection automatic)
3 ½ digit liquid crystal display with 0.5” digits
LCD “LO BAT” legend to indicate low battery
Environmental
Operating Temperature: 0 °C to 50 °C, 75 % RH, Non-condensing
Storage Temperature: -40 °C to 71 °C, 0-100 % RH, Non-condensing
Power and Physical:
Power Standard: alkaline 9 V battery
Size: 3.5” W x 7.0” H x 1.2” D
Weight: 0.6 lb (including battery)
Included Accessories:
Protective carrying pouch
Two 40 inch test leads with probes and insulated, detachable alligator clips
Operation and Maintenance manual