直接联系认证专家

SPECIFICATIONS
Relative Humidity Range: 10 to 95 %RH, 10 to 40 L/min
Relative Humidity Range: 95 to 98 %RH, 10 to 20 L/min
Frost Point Temperature Range: –33.60 to 0 °C
Dew Point Temperature Range: –36.94 to 70 °C
Chamber Fluid Temperature Range: –10 to 72 °C
Chamber Fluid Temperature Control Stability: 0.002 °C
Chamber Temperature Uniformity: < 0.04 °C
Chamber Fluid Temperature Heating Rate: from –10 to 72 °C: 0.5 °C per Minute (average)
Chamber Fluid Temperature Cooling Rate: from 72 to 0 °C: 0.5 °C per Minute (average)
Chamber Fluid Temperature Cooling Rate: from 0 to –10 °C: 0.2 °C per Minute (average)
Display Resolution: 0.001
Temperature Specification: 0.027 °C
Gas Type: Air or Nitrogen
Gas Pressure Rating: (MAWP) 175 psiG
Gas Flow Rate Range: 10 to 50 L/min
Gas Flow Rate Specification: 5% of full scale
Saturation Pressure Range: Ambient to 160 psiA
Saturation Pressure Specification: 0.02% of full scale
Test Chamber Pressure Range: Ambient
Supply Pressure Range: Ambient to 150 psiG
Supply Pressure Specification: ±1 psiG
Test Chamber Dimensions: 12" H x 12" W x 10" D (30.48 cm x 30.48 cm x 25.4 cm)
Physical Dimensions: 22" H x 36" W x 23" D (55.88 cm x 91.44 cm x 58.42 cm)
Access Port: 2 @ Ø1.875" (47.6 mm) located on right side
UNCERTAINTY
RH Uncertainty: 10 to 95 %RH, 0 to 70 °C, 10 to 40 L/min: 0.5% of reading
RH Uncertainty: 95 to 98 %RH, 0 to 70 °C, 10 to 20 L/min: 0.5% of reading
Dew/Frost Point Uncertainty: < 0 °C, 10 to 40 L/min: 0.05 °C
Dew Point Uncertainty: 0 to 70 °C, 10 to 40 L/min: 0.08 °C
Test Chamber Temperature Uncertainty: -10 to 72 °C: 0.031 °C
Test Chamber Pressure Uncertainty: Ambient: 0.007 psiA
UTILITIES
Electrical Power: 200-230/210-240 V~, 10 A, 1 Ø, 50/60 Hz
Gas Supply: 175 psiG @ 50 L/min
ENVIRONMENTAL
Operating Temperature: 15 to 30 °C
Storage Temperature: 0 to 50 °C
Humidity: 5 to 95% Non-condensing