Specimen clamp: clamp type A specimen 15 pcs, clamp type B specimen 5 pcs
Striking device: Type A and Type B.
Impact speed: 2000±200mm/s
Specimen dimension: A: Length: 20.0±0.25mm, Width: 2.5±0.1mm,Thickness: 2.0±0.1mm
: B:Length: 31.75±6.35mm,Width: 6.35±0.51mm,Thickness: 1.91±0.13mm
Size of cooling working chamber: 180×570×150(mm)
Temperature range: -70℃- 0℃/ -80℃-0℃
Temperature fluctuation: ±0.5℃
Temperature deviation: ≤±0.5℃
Temperature uniformity: ±0.5℃
Cooling speed: To lowest temperature need about 2 hours
Refrigerant: R404A
Power Supply: 220V, 50HZ/60HZ, 4000W
Product Description:
Low Temperature Brittleness Tester is used to determine the brittleness temperature, specimens are secured to a specimen holder using a torque wrench. The specimen holder is immersed in a bath containing a heat-transfer medium, which has been cooled. The specimens are struck at a specified linear speed and then examined for failures. The brittleness temperature is defined as the temperature at which 50% of the specimens would probably fail.
Test Standards: ISO 974
Product Features:
Striking device and its accessories is made of XAR450 Germany anti-abrasion steel plate.
Thermal insulation material: use high density glass fiber cotton, insulation thickness of 80mm, ensure the temperature is constant during the testing .
Has mixing system to let temperature in low temperature chamber keep uniformity.
Has feet wheel for moving machine easily.
Heating system use stainless steel heating tube.
The cooling compressor use Italian brand aspera.
PID Temperature controller to control temperature accurately.