For full functionality of this site it is necessary to enable JavaScript.

Additel 875-155干井低温校准器(-40°Cđến155°C)

Specification

Temperature Range at 23℃: -40°C to 155°C

Display Accuracy: ±0.18°C at Full Range

Stability (30 min): ±0.01°C at Full Range

Axial Uniformity at 60 mm (2.4 in): ±0.07°C at Full Range

Radial Uniformity: ±0.01°C at Full Range

Loading Effect:

±0.1°C (Display Sensor)

±0.02°C (External Sensor)

Hysteresis (Display Sensor): 0.025°C

Environmental Conditions:

8°C to 38°C guaranteed accuracy 

0°C to 50°C, 0% to 90% RH non-condensing, 3000 M altitude for normal operation

Storage Conditions: -20°C to 60°C

IP Rating: IP20

Immersion Depth: 150 mm (5.9 in)

Insert OD: 25.8 mm (1.02 in)

Heating Time:

13 min: -40°C to 155°C

5 min: -40°C to 23°C

8 min: 23°C to 155°C

Cooling Time:

28 min: 155°C to -40°C

8 min: 155°C to 23°C

20 min: 23°C to -40°C

Typical Time to Stability: 10 min

Resolution: 0.01°C

Units: °C, °F, and K

Display: 6.5 in (165 mm) color touch screen

Size (H x W x D): 320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in)

Weight: 9.9 kg (21.8 lbs)

Power Requirements: 90-254 VAC, 45-65 Hz, 580 W

Mechanical Testing:

Vibration: 2 g (10-500 Hz), 30 min for 2 sides

Impact: 4 g three times

Drop test: 500 mm (19.6 in)

Communication: USB A, USB B, RJ45, WiFi, Bluetooth

Localization: English, Chinese, Japanese, Russian, German, French, Italian, and Spanish


Details
  • 质量承诺
  • 正品保修
  • 送货到家
  • 交易简单化

注册收新闻 - 获得优惠活动的机会