For full functionality of this site it is necessary to enable JavaScript.

ESCO CBI-III-2G8 Containment Barrier Isolator - III (1.2 m)

Main Chamber Nominal Size (Width): 1.2 meters 

Working Chamber Dimensions - Min (W x D x H): 1.2 m x 0.55 m x 0.77 m 

Working Chamber Dimensions - Max (W x D x H): 1.2 m x 0.63 m x 0.84 m 

External Dimension (with one Pass Chamber) (W x D x H): 

With Adjustable Base Stand (Min): 1.82 m x 0.92 m x 2.55 m  

With Adjustable Base Stand (Max): 1.82 m x 0.92 m x 2.85 m 

External Dimension - Small Pass-through Chamber (W x D x H): 0.60 m x 0.52 m x 0.76 m 

Glove Port Height (Min): 1000 mm 

Glove Port Height (Max): 1300 mm 

Chamber Environment: ISO Class 5 all chambers (Grade A) 

Pre-filter: 

Process Chamber: G4, polyester media 

Pass-through Chamber: G4, polyester media 

Filter Type - Chamber Inlet: HEPA (H14) Filter with Integral Mesh Guard and Gasket Seal 

Filter Efficiency - Chamber Inlet: 99.995% MPPS 

Filter Type - Chamber Exhaust: HEPA (H14) Filter with Integral Mesh Guard and Gasket Seal 

Filter Efficiency - Chamber Return and Exhaust: 99.995% MPPS 

Lighting Level: ≥ 500 Lux (6000 K) 

Sound Level: ≤ 80 dBA 

Airflow Type: Unidirectional/ Laminar Airflow 

Chamber Operating Pressure: Negative 125 Pa 

Isolator Construction: 

Chamber: SS 316L 

Service Housing: SS 304 

Support Frame: SS 304 

Isolator Finish: 

Chamber Internal: ≤ 0.4 Ra 

Chamber External: ≤ 0.6 Ra 

Service Housing External: ≤ 0.6 Ra 

Support Frame: ≤ 1.0 Ra 

Electrical Requirements: 220-240 VAC, 50/60 Hz, 1Ø 

Compressed Air Requirement (by Client): 

If no on-board compressor: 2 Bar-g Pressure at 5 L/sec ✔  

Exhaust Duct Requirements (by Client) unless Integral Catalytic Converter is Included: 10“ Duct from Isolator to Outside 

Air Change per Hour: 1570 ACH 

Volumetric Flow Rate: 916 cmh 

Downflow Velocity: 0.4 m/s  

Net Weight: TBD