DC mV: 600.0 mV/ ± 0.5%
DC V: 6.000 V,  60.00 V, 600.00 V / 0.5%
AC mV(True RMS): 600.0 mV/ 1.0 % 
ACV (True RMS): 6.000 V, 60.00 V, 600.0 V /  1.0 %
Continuity: yes
Ohms: 600.0 Ω, 6.000 kΩ, 60.00 kΩ, 600.0 kΩ, 6.000 MΩ /  0.9 %, 40.00 MΩ / 5 %
Diode Test: yes           
C: 1000 nF, 10.00 µF, 100.0 µF, 9999 µF
Lo-Z C: 1 nF to 500 µF / 10%
ACA (True RMS) : 6.000 A, 10.00 A / 1.5% 
DCA: 6.000 A,10.00 A / 1.0%  
Hz: 99.99 Hz, 999.9 Hz, 9.999 kHz, 50 kHz /  0.1%
Size: 167 x 84 x 46 mm (6.57" x 3.31" x 1.82")
Weight: 550 g
Includes: Holster, Installed 9 V battery,Test leads, User manual
 Details
| Maximum voltage between any terminal and earth ground |  | 
| Surge protection | 
|  | 6 kV peak per IEC 61010-1 600 VCAT III, Pollution Degree 2 |  | 
| Display | 
| Digital: | 6,000 counts, updates 4 per second |  | 
| Bar graph | 
|  | 33 segments, updates 32 per second |  | 
| Operating Temperature |  | 
| Storage Temperature |  | 
| Battery type | 
|  | 9 volt Alkaline, NEDA 1604A/ IEC 6LR61 |  | 
| Battery Life | 
|  | 400 hours typical, without backlight |  | 
| Accuracy Specifications | 
| DC millivolts | 
| Range: | 600.0 mV |  
| Resolution: | 0.1 mV |  
| Accuracy: | ± ([% of reading] + [counts]): 0.5% + 2 |  | 
| DC volts | 
| Range/Resolution: | 6.000 V / 0.001 V |  
| Range/Resolution: | 60.00 V / 0.01 V |  
| Range/Resolution: | 600.00 V / 0.1 V |  
| Accuracy: | ± ([% of reading] + [counts]): 0.5% + 2 |  | 
| AC millivolts1 true-rms | 
| Range: | 600.0 mV |  
| Resolution: | 0.1 mV |  
| Accuracy: | 1.0 % + 3 (dc, 45 Hz to 500 Hz)2.0 % + 3 (500 Hz to 1 kHz) |  | 
| AC volts1 true-rms | 
| Range/Resolution: | 6.000 V / 0.001 V |  
| Range/Resolution: | 60.00 V / 0.01 V |  
| Range/Resolution: | 600.0 V / 0.1 V |  
| Accuracy: | 1.0 % + 3 (dc, 45 Hz to 500 Hz)2.0 % + 3 (500 Hz to 1 kHz) |  | 
| Continuity | 
| Range: | 600 Ω |  
| Resolution: | 1 Ω |  
| Accuracy: | Beeper on < 20 Ω, off > 250 Ω; detectsopens or shorts of 500 μs or longer. |  | 
| Ohms | 
| Range/Resolution: | 600.0 Ω / 0.1 Ω |  
| Range/Resolution: | 6.000 kΩ / 0.001 kΩ |  
| Range/Resolution: | 60.00 kΩ / 0.01 kΩ |  
| Range/Resolution: | 600.0 kΩ / 0.1 kΩ |  
| Range/Resolution: | 6.000 MΩ / 0.001 MΩ |  
| Accuracy: | 0.9 % + 1 |  
| Range/Resolution: | 40.00 MΩ / 0.01 MΩ |  
| Accuracy: | 5 % + 2 |  | 
| Diode Test | 
| Range/Resolution: | 2.00 V / 0.001 V |  
| Accuracy: | 0.9% + 2 |  | 
| Capacitance | 
| Range/Resolution: | 1000 nF / 1 nF |  
| Range/Resolution: | 10.00 µF / 0.01 µF |  
| Range/Resolution: | 100.0 µF / 0.1 µF |  
| Range/Resolution: | 9999 µF / 1 µF |  
| Range/Resolution: | 100 µF to 1000 µF |  
| Accuracy: | 1.9% + 2 |  
| Range/Resolution: | > 1000 µF |  
| Accuracy: | 5% + 20% |  | 
| Lo-Z Capacitance | 
| Range: | 1 nF to 500 µF |  
| Accuracy: | 10% + 2 typical |  | 
| AC amps true-rms (45 Hz to 500Hz) | 
| Range/Resolution: | 6.000 A / 0.001 A |  
| Range/Resolution: | 10.00 A / 0.01 A |  
| Accuracy: | 1.5% + 3 |  
|  | 20 A overload for 30-seconds max |  | 
| DC amps | 
| Range/Resolution: | 6.000 A / 0.001 A |  
| Range/Resolution: | 10.00 A / 0.01 A |  
| Accuracy: | 1.0% + 3 |  
|  | 20 A overload for 30-seconds max |  | 
| Hz (V or A input)2 | 
| Range/Resolution: | 99.99 Hz / 0.01 Hz |  
| Range/Resolution: | 999.9 Hz / 0.01 Hz |  
| Range/Resolution: | 9.999 kHz / 0.001 kHz |  
| Range/Resolution: | 50 kHz / 0.01 kHz |  
| Accuracy: | 0.1% + 2 |  | 
Notes:
 1) All ac voltage ranges except Auto-V/LoZ are specified from 1 % to 100 % of range. Auto-V/LoZ is specified from 0.0 V.
 2) Temperature uncertainty (accuracy) does not include the error of the thermocouple probe.
 3) Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current.
 4) Temperature Range/Resolution: -40 °F to 752 °F / 0.2 °F 
 2) Frequency is ac coupled, 5Hz to 30 kHz for ac voltage
| Mechanical & General Specifications | 
| Size | 
|  | 167 x 84 x 46 mm (6.57" x 3.31" x 1.82") |  | 
| Weight |  | 
Brochure »
Data Sheet »