Gastec 109A Airtec(压缩空气)油雾检测管(0.3~1.5 mg / m3)

Indicator: 0.3 ~ 1.5 mg/m3
Measuring Range: 0.3 ~ 1.5 mg/m3
Sampling Rate: 1000mL/min (60000mL)
Sampling Time:  60 minute
Colour Change: White → Dark green 
Reaction Principle: Oil + I2O5 + H2S2O7 → I2
Coefficient of Variation: 15% (for 0.3 to 0.5 mg/m3) , 10% (for 0.5 to 1.5 mg/m3)
Nr. of tests: 10

Details

Clatalog

 

Performance
Measuring Range 0.3 to 1.5 mg/m3
Sampling Rate 1000mL/min (60000mL)
Correction Factor 1
Sampling Time 60 minutes
Colour Change White → Dark green
Reaction Principle Oil + I2O5 + H2S2O7 → I2
Coefficient of Variation 15% (for 0.3 to 0.5 mg/m3) , 10% (for 0.5 to 1.5 mg/m3)
Shelf Life 2 Years
Corrections for temperature Unnecessary for 0 to 40℃
Corrections for humidity Unnecessary for R.H.0 to 90%
Store the tubes at cool and dark place.
Calibration gas generation Diffusion tube method

 

Airtec tube - Compressed breathing air detector tubes

Airtec Tube for Cylinder / Compressed Breathing Air / Air Line Test.
Airtec tubes are a convenient and simple system for testing of a gas cylinder, compressed breathing air and air line. Industrial operations often produce, or are performed in the presence of harmful airborne contaminants. When self contained breathing apparatus or other devices are used for respiratory protection, the quality of the breathing air requires special attention. Contaminants entering the compressor or contaminants generated by the compressor or cylinder can be harmful to the worker and the respiratory equipment.
 




Measurement procedure:


1.Install a reducer pressure with gauge and flow metre to a cylinder, compressor or air line and adjust the flow metre the required setting.
  2.Break tips off a fresh detector tube in the tube tip breaker and insert a tube into a tube holder.
 
3.Attach the rubber tube holder to the flow metre outlet. Make certain the tube arrowon the tube pointing in a down direction.
  4.Turn on the cylinder or compressor and confirm the flow metre according to each Airtec tube specifications.
 
5.Time the flow rate with a stopwatch.
  6.As soon as sampling time has elapsed turn off the cylinder or compressor, and remove the tube from the tube holder and then read the colour change layer immediately.

 