For full functionality of this site it is necessary to enable JavaScript.

Gastec 1A Airtec(压缩空气)一氧化碳探测器管 CO(5~50 ppm)

Indicator: 5 to50 ppm
Measuring Range: 5 to50 ppm
Sampling Rate: 100mL/min(300mL)
Sampling Time: 3 minute
Colour Change: Yellow → Dark brown
Reaction Principle: CO + Na2Pd(SO3)2 → Pd + CO2 + SO2+Na2SO3
Coefficient of Variation: 10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm)
Nr. of tests: 10

Details

Catalog

Performance

Measuring Range 5 to50 ppm
Sampling Rate 100mL/min(300mL)
Correction Factor 1
Sampling Time 3 minutes
Colour Change Yellow → Dark brown
Reaction Principle CO + Na2Pd(SO3)2 → Pd + CO2 + SO2+Na2SO3
Coefficient of Variation 10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm)
Shelf Life 2 Years
Corrections for temperature Unnecessary for 0 to 40℃
Corrections for humidity Unnecessary for R.H.0 to 90%
Store the tubes at cool and dark place.
Calibration gas generation High pressure gas cylinder method
TLV-TWA TLV-STEL Explosive range
25 ppm - 12.5 to 74%

 

Airtec tube - Compressed breathing air detector tubes

Airtec Tube for Cylinder / Compressed Breathing Air / Air Line Test.
Airtec tubes are a convenient and simple system for testing of a gas cylinder, compressed breathing air and air line. Industrial operations often produce, or are performed in the presence of harmful airborne contaminants. When self contained breathing apparatus or other devices are used for respiratory protection, the quality of the breathing air requires special attention. Contaminants entering the compressor or contaminants generated by the compressor or cylinder can be harmful to the worker and the respiratory equipment.
 




Measurement procedure:


1.Install a reducer pressure with gauge and flow metre to a cylinder, compressor or air line and adjust the flow metre the required setting.
  2.Break tips off a fresh detector tube in the tube tip breaker and insert a tube into a tube holder.
 
3.Attach the rubber tube holder to the flow metre outlet. Make certain the tube arrowon the tube pointing in a down direction.
  4.Turn on the cylinder or compressor and confirm the flow metre according to each Airtec tube specifications.
 
5.Time the flow rate with a stopwatch.
  6.As soon as sampling time has elapsed turn off the cylinder or compressor, and remove the tube from the tube holder and then read the colour change layer immediately.

 

  • 质量承诺
  • 正品保修
  • 送货到家
  • 交易简单化

注册收新闻 - 获得优惠活动的机会