Gastec 221LL 氯离子检测管(Cl-)(10~200 mg / L)
制造商: Gastec Model: 221LL - 联系
 
Indicator: 10 ~ 200 mg/L
Measuring Range: 10 ~ 200 mg/L
Sampling Time:  
Colour Change: 
Reaction Principle: 
Nr. of tests: 10
| Measuring Range | 10 to 200 mg/L | 
| Sampling Time | 1 minutes | 
| Detecting Limit | 3 mg/L | 
| Colour Change | Brown → White | 
| Reaction Principle | 2Cl- + Ag2CrO4 → 2AgCl + CrO42- | 
| Shelf Life | 3 Years | 
| Corrections for temperature & humidity | Unnecessary | 
| Store the tubes at cool and dark place. | |
| Substance | Concentration | Interference | Changes colour by itself to | 
| CN- | ≧ 10 mg/L | + | White ( ≧ 1mg/L) | 
| Br- | ≧ 10 mg/L | + | White ( ≧ 10mg/L) | 
| SCN- | ≧ 1 mg/L | Unreadable | White ( ≧ 100mg/L) | 
| Fe2+ | ≧ 20 mg/L | No ( ≦ 1000mg/L) | |
| Fe3+ | ≧ 20 mg/L | + | White ( ≧ 20mg/L) | 
| F- | ≦ 1000 mg/L | No | No( ≦ 1000mg/L) | 
| I- | ≧ 20 mg/L | + | Pale yellow ( ≧ 20mg/L) | 
| S2- | ≧ 3 mg/L | + | Blackish gray ( ≧ 3mg/L) | 
| SO42- | ≦ 1000 mg/L | No | No ( ≦ 1000mg/L) | 
| Substance | Correction Factor | Measuring Range | 
| Br- | 2.4 | 24 to 480 mg/l | 
(i.e. in solution).
| 
 Dissolved sulphide detector tube 
211H / 211M / 211 / 211LL 
![]()  | 
 Liquid detector tube 
  | 
* Adjustment for pH is required when measuring with liquid detector tubes.