Gastec 91PL 甲醛气体检测管 HCHO (0.01-0.80 ppm)
制造商: Gastec Model: 91PL - 报价要求
- 联系
Calibration Range: 0.01 -0.20 ppm
Measuring Range: 0.01-0.80 ppm
Sampling (flow) Rate: 200mL/min
Sampling Volume: 6000mL, 2000mL
Sampling Time: 30min, 10min.
Colour Change: Pale yellow → Pink
Nr. of tests: 10
Automatic Gas Sampler: GSP-300FT-2
Calibration Range | (0.01)-0.20 ppm |
Measuring range | 0.01-0.80 ppm |
Sampling (flow) Rate | 200mL/min |
Sampling Volume | 6000mL, 2000mL |
Sampling Time | 30min, 10min. |
Detecting Limit | 0.005 ppm |
Colour Change | Pale yellow → Pink |
Corrections for temperature/humidity | Temperature |
Shelf life | 1 year (in the refrigerator) |
Accuracy | CV=5-10% (CV: Coefficient of variation = σ: standard deviation / mean value x 100) |
Substance(s) | Interference | Colour change by itself |
Ammonia | max. 2 ppm | No | No |
Nitrogen dioxide | max. 1 ppm | No | No |
Acetaldehyde | "+" error (higher reading) |
produces pinkish discolouration | |
Acetone | max. 0.3 ppm | No | Produces pinkish stain around tube opening (inlet) |
Ethanol | No | No | |
Ethyl acetate | No | No | |
p-dichlorobenzene | No | No | |
Toluene | No | No | |
Scrubber 1 (black):removes acetone Scrubber 2 (purple): Ammonia and nitrogen dioxide are removed. The presence of ammonia causes the purple scrubber 2 to turn yellow. The presence of nitrogen dioxide causes scrubber 2 to turn a darker shade of purple |
Calibration gas generation | Diffusion tube method (analysis: by absorptiometry, acetylacetone method) |
Electric sampling pump This detector tube is suited for measuring formaldehyde in indoor environments. The GSP-200 / GSP-300FT series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance. A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value. Formaldehyde indoor concentration threshold: 0.08 ppm (mean value during 30 minutes) |
TLV-TWA | TLV-STEL | Explosive range |
-- -- | C0.3ppm(2014) | 7.0 to 73% |
- 质量承诺
- 正品保修
- 送货到家